Specifications 

Technical Data

 

Measuring range 

1.000 kg 

Resolution 

200 g 

Minimum load 

10 kg 

Measurement error 

± 1% of the measuring range  

Tare range 

100 % F.s. 

Response time 

≥ 1 second 

Safety load 

150 % (On no account may this be exceeded) 

Display 

LCD with background lighting 

Digit height 

20 mm 

Ambient temperature 

-10°C 

 +40°C 

Power supply 

3 x 1,5 V AA-Battery (operating time up to 65 h) 

Weight 

approximately 1,4 kg including batteries / hook and 

shackles  

 

  

 

 

 

 
 
 

Model 

 

 

 

 

PCE-CS 1000N 

 

20mm 

 

43mm 

 

27mm 

 

230mm 

 

Содержание PCE-CS 1000 N

Страница 1: ...US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire S...

Страница 2: ...c switch off 10 6 2 br 3 Set background lighting 10 6 3 IdL Lighting duration 10 7 Troubleshooting 10 8 Custom settings 11 8 1 Equipment configuration 11 8 1 1 Change resolution 11 8 1 2 Automatic zer...

Страница 3: ...the manufacturer and the operator adhere to certain requirements When using the measuring equipment no risk must be posed to the operator The operator or the person responsible for operating the measu...

Страница 4: ...s to be allowed to be dropped or be pulled along the ground or pulled at an angle The maximum load as stipulated in the operating instructions MAX must never be exceeded this also includes the maximum...

Страница 5: ...functioning correctly After every 12 months or after 50 000 measurements depending on which comes first Check all load bearing parts for example hairline cracks etc Only trained personnel are to carr...

Страница 6: ...nge Tare range 100 F s Response time 1 second Safety load 150 On no account may this be exceeded Display LCD with background lighting Digit height 20 mm Ambient temperature 10 C 40 C Power supply 3 x...

Страница 7: ...ontrol 1 x operating instructions 4 Equipment description 4 1 Measuring equipment 4 1 Power button 4 2 Infrared sensor 4 3 Display 4 4 Tare Zero button Display Shows the measured values Power button T...

Страница 8: ...nts 4 8 Left 4 9 Setup Upwards Zero Move one value upwards Set the measuring value to zero Hold button Enter Saves the value Confirms the entry Downwards Moves one position to the right Unit change Ch...

Страница 9: ...h the remote control 5 2 Zero Press the button until an acoustic signal sounds Z appears on the display The remote control can also be used to do this Press 0 Make sure the value is stable Otherwise E...

Страница 10: ...ntil End appears on the display 6 3 IdL Lighting duration To set the lighting duration press three times ldl lighting duration appears on the display Press the T to set the lighting You can set it on...

Страница 11: ...until End appears on the display 8 1 2 Automatic zero tolerance To set the automatic zero tolerance press twice A percent value is displayed The tolerance can be changed with the arrow buttons on the...

Страница 12: ...ess until End appears on the display 8 1 9 Setting gravity To set the gravity press nine times value is displayed The value can be changed with the up and down buttons The lowest value is 0 000 and th...

Страница 13: ...and for the 3 point calibration for example 1 3 2 3 or the full load The individual adjustment points can be freely set on the scale and can be adapted to the weights provided by the operator The cod...

Страница 14: ...Press again LoAd2 appears on the display If this value is sufficient then switch the equipment off or press to continue Press repeatedly 00000 appears on the display The second measuring unit can be...

Страница 15: ...h ElektroG R cknahme und Entsorgung von Elektro und Elektronikaltger ten our equipment can be returned We will either re use the equipment or it will be disposed of by a recycling company according to...

Отзывы: