1  Introduction 

Thank you for purchasing an infrared thermometer PCE-889B from PCE Instruments. 

2  Safety notes 

Please read this manual carefully and completely before you use the device for the first time. The device 
may only be used by qualified personnel and repaired by PCE Instruments personnel. There is no 
warranty of damage or injuries caused by non-observance of the manual.

 

 
- The device may only be used in the approved temperature range. 

The case should only be opened by qualified personnel of PCE Instruments. 

- Only use PCE replacement parts or appropriate equivalent.

 

- The instrument should never be placed with the user interface facing an object (e.g. keyboard side on a 
  table). 
- You must not make any technical changes to the device. 
- The appliance should periodically be cleaned with a dry cloth / use only pH-neutral cleaner, no 
  abrasives or solvents.

 

- Be extremely careful when the laser beam is turned on. 

Do not let the beam enter your eye, another person’s eye or the eye of an animal. 

- Be careful not to let the beam strike your eye via a reflective surface. 
- Keep the instrument away from explosive gases. 
- It is not recommended to measure shiny or polished metal surfaces (stainless steel, aluminum, etc.) with 
  this thermometer, see information on emissivity. 

-

 The device cannot measure through transparent surfaces such as glass. It will measure the surface 

  temperature of the glass instead. 
- Steam, dust, smoke, etc., can obstruct the meter

’s optics and thus affect measurement accuracy. 

 
This user's handbook is published by PCE Instruments without any guarantee. 
 
We expressly point to our general guarantee terms which can be found in our general terms of business. 
 
If you have any questions please contact PCE Instruments. 
 

3  Specifications 

Measurement ranges

 

-50 ... +1000 °C

 

Resolution

 

0.1 °C

 

Accuracy

 

±1 %

 

Response time

 

<150 ms

 

Distance:spot ratio

 

30:1

 

Adjustable emissivity, depending on material

 

0.1 ... 1.0

 

Spectral range

 

8 ... 14 µm

 

Visible laser beam for targeting

 

dual laser beam (class 2)

 

Dimensions

 

146 x 104 x 43 mm

 

Weight

 

163 g

  

Temperature units

 

°C / °F

  

Display functions

  

HOLD function, Auto Power Off, MAX value 
indication

 

Backlight

 

 

Indication when temperature is exceeded

 

 

Continuous measurement

 

 

Alarm output

 

 

Power supply

 

9 V battery

 

Содержание PCE-889B

Страница 1: ...ruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce i...

Страница 2: ...tents 1 Introduction 3 2 Safety notes 3 3 Specifications 3 4 System description 4 5 Background information 5 6 Button functions Instructions 6 7 Disposal 8 8 Contact 8 8 1 PCE Instruments UK 8 8 2 PCE...

Страница 3: ...not to let the beam strike your eye via a reflective surface Keep the instrument away from explosive gases It is not recommended to measure shiny or polished metal surfaces stainless steel aluminum et...

Страница 4: ...artment cover Handle Display Data hold Laser on symbol Lock symbol High alarm and low alarm symbols C F symbol Low battery indicator Emissivity symbol and value MAX temperature value MAX temperature s...

Страница 5: ...urfaces have an emissivity of 0 95 pre set in the device Inaccurate readings will result from measuring shiny or polished metal surfaces To compensate cover the surface to be measured with masking tap...

Страница 6: ...the display press the up and down buttons to adjust the desired values The mode structure is as follows Emissivity EMS adjustment The Emissivity EMS is digitally adjustable from 0 10 to 1 0 by means...

Страница 7: ...s the device is locked to stay on Note The meter automatically compensates for deviations in the ambient temperature Please bear in mind that it will take up to 30 minutes to adapt to a new ambient te...

Страница 8: ...ur devices back We either re use them or give them to a recycling company which disposes of the devices in line with law 8 Contact If you have any questions about our range of products or measuring in...

Отзывы: