Code Table:
SWELL
DIP
OUTAGE
COMMENT
CODE
1
2
4
Codes can be added together
DISPLAY FORMAT:
First column: sequential number of events.
Second column: elapsed time since the time of start. The elapsed time format is
(DAYS, HOURS, MINUTES). The maximum time is 99 day 24 hours 60 minutes.
The third column: number of cycles.
The fourth column: code of transient events. There might be more than one
transient condition occurs in one event.
NOTE:
In the mode of TRANSIENT CAPTURE, the analyzer takes 128 samples of
each cycle for each phase continuously.
NOTE:
When the users press the TRANSIENT button to review the recorded
events, the analyzer stops capture operation until the TRANSIENT is pressed again.
The timer is also stopped when users press the TRANSIENT button. So the time
stamp is not correct when users press the TRANSIENT button to resume operation.
NOTE:
The analyzer can record up to 28 events. When the analyzer has recorded
28 events, it stops the capture operation, turns the backlight on, and displays the 28
transient events.
NOTE:
The code can be added together to indicate two or three conditions. For
example, if code is 6, that means DIP and OUTAGE (2+4).
NOTE:
The longest duration for capture operation is 99 days. So users should use
external 12V DC power adaptor for capture operation.
WARNING:
Users must select 50 or 60 Hz for transient capture. If users select
AUTO for frequency, the unit will not allow users to enter the TRANSIENT capture,
and will beep to warn users.