PCE-MSM 4, PCE-322A, PCE-323 Sound Level Meter 

User Manual 

User manuals in various languages (français, 

italiano, español, português, nederlands, türk, polski, 

русский, 

中文

) can be found by using our 

product search on: www.pce-instruments.com 

Last change: 7 November 2022 

v1.0 

E

nglis

h

 

Содержание PCE-322A

Страница 1: ...E 323 Sound Level Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Las...

Страница 2: ...and off 6 5 1 Making a measurement 6 6 Automatic power off 9 7 Data logging PCE 322A PCE 323 only 10 7 1 Clearing the memory 10 7 2 Setting the memory interval 10 8 Replacing the battery 10 9 Bluetoo...

Страница 3: ...ure Do not expose the device to shocks or strong vibrations The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any...

Страница 4: ...1 k Data interface USB Bluetooth 4 0 PCE 323 only Bracket 3 8 connection Functions MIN MAX HOLD display when exceeding or falling below the sound level Automatic power off after approx 15 minutes of i...

Страница 5: ...truments 3 4 Device description No Description 1 Microphone with windscreen 2 Connections for power supply mini USB analogue outputs calibration screw 3 Display 4 Keyboard 5 Battery compartment rear 1...

Страница 6: ...ic power off and to start the Bluetooth connection PCE 323 only 3 Switching the backlight on and off 4 FAST SLOW key to switch between fast and slow time rating 5 A C key for switching between frequen...

Страница 7: ...t measured value 5 Time rating Fast fast 125 ms 6 Time rating Slow slow 1 s 7 Current measured value is above the set measurement range 8 Measurement range scale 9 Date and time 10 Measured value 11 A...

Страница 8: ...le If a range is exceeded or undercut this is indicated by Over or Under 5 1 3 Setting the time rating To set the time rating press the FAST SLOW key With the Fast setting the time weighting is 125 ms...

Страница 9: ...y and switch on the meter As soon as TIME is flashing you can release the SETUP key The currently set date is now displayed You can then select the minute hour day month and year one after the other u...

Страница 10: ...PCE Instruments 8 Setting the hour Setting the day Setting the month...

Страница 11: ...and time To do this go one function further rSt Reset is displayed for resetting Now press the HOLD key and the date and time will be reset Now you can set the date and time 6 Automatic power off To...

Страница 12: ...eter Press and hold the backlight key and switch the meter back on As soon as Int is displayed you can release the backlight key Use the LEVEL key to adjust the displayed value A memory interval betwe...

Страница 13: ...to activate the Bluetooth function on your PCE 323 Then open the app and connect to the meter via the app Select Environment Meter as the instrument type Note The connection must not be established v...

Страница 14: ...on it 3 Then click on Install to install the driver 4 Restart the PC after the driver installation has been completed 5 Once the PC has restarted you can switch on the meter and connect it to a free...

Страница 15: ...P210x USB to UART Bridge COMX then to Port Settings and Advanced Port Settings 7 Now start the software installation by executing the Setup exe in the root directory and following the information on t...

Страница 16: ...ation on the version of the software 10 2 2 Real time measurement Start Time Time from the start of the measurement MIN Lowest volume of the measurement series with point in time MAX Highest volume of...

Страница 17: ...and B Minimum determined value between A and B Cursor B Value cursor B Average Between A and B Measured average value between A and B Quantity Between A and B Measured values between A and B You can...

Страница 18: ...measurement Stop Stop real time measurement Clear Data Clear data Setup Setting the sampling rate and maximum measured values Read out the internal data logger of the meter Manual Select communication...

Страница 19: ...C Output voltage 10 mV dB Resistor 1 k 13 Calibration To perform a calibration connect the meter to a class 2 or better sound calibrator Set the meter as follows Frequency rating A Time rating Almost...

Страница 20: ...ies Due to the contained pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012...