15

 

5.6 Data logging 

At most one hundred files (F00-F99, one group as one file) can be stored inside the gauge. By 

simply pressing the   key after a new measurement finishes-the screen showing the “AVE” icon, the 
measured hardness/strength group values will be saved to memory. The new saved file is appended as 
the last file of the memory. This function provides the user with the ability to view/delete a file/group 
previously saved in memory. 

5.6.1 Viewing stored file/Group 

To view the memory data, follow the steps: 

1)

 

Press the   key to activate the data logging function. The memory icon will appear. It will display 
the current file name, the test parameter of the group data and the mean value of the group. If there 
is no data in the memory, it will display: <E04>, which means no memory data, and then return 
back. 

2)

 

Use the 

 key and the 

 key to select the desired file to view. 

3)

 

Press the   key to see details of that group data.  

4)

 

Use the 

 key and the 

 key to view each single measured data in that group while viewing 

details. 

5)

 

Press the 

 key to return to previous screen at any time during data logging. 

5.6.2 Deleting selected file/Group 

The user may require deleting a file from the instrument memory. The procedure is outlined in 

the following steps. 
1)

 

Press the   key to activate the data logging function. The memory icon will appear. It will display 
the current file name, the test parameter of the group data and the mean value of the group. If there 
is no data in the memory, it will display: <E04>, which means no memory data, and then return 
back. 

2)

 

Use the 

 key and the 

 key to scroll to the file that will be deleted. 

3)

 

Press the   key on the desired file. It will automatically delete the file, and display “-DEL”. 

4)

 

Press the 

 key, at any time, to exit the data logging function and return to measurement mode. 

Note

 Do not shut down the instrument while deleting data. It could lead to unpredicted 

consequence if shutting down while deleting.  

5.7 Print Function 

At the end of the inspection process, or end of the day, the user may require the readings be 

printed. You can print out the measurement result immediately after each testing process, by easily 
pressing the  key.  If you want to print the data stored in the instrument memory, then following is 
the steps: 
1.

 

Press the   key to activate the data logging function. The memory icon will appear. 

2.

 

Use the 

 key and the 

 key to select the desired file. 

3.

 

Press the   key to print the selected file. This operation will print out all the data in current file. 

4.

 

Press the 

 key to exit the data logging functions and return to measurement mode. 

Note: 

z

 

Printing function is unavailable while charging. 

z

 

Printing can be stopped by pressing the

 key. 

z

 

Do not open the cover of the paper compartment during printing. Otherwise the instrument may 
not print normally. 

z

 

Over high ambient humidity (above 85% of relative humidity) or over low ambient humidity 
(below 20% of relative humidity) may reduce the print quality. 

z

 

Printing with paper that has been stored for over long period of time or of poor quality may 

Содержание PCE-2800

Страница 1: ...2 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kin...

Страница 2: ...ness Testing Principle 9 3 Preparation 9 3 1 Instrument Preparation and Inspection 9 3 2 Impact Device Selection 9 3 3 Preparation of the Sample Surface 10 4 Testing Program 11 4 1 Start Up 11 4 2 Loa...

Страница 3: ...15 Error Code Reference 17 6 Maintenance Servicing 17 6 1 Impact Device Maintenance 17 6 2 Instrument Maintenance Program 18 6 3 Fault Analysis Evacuation 18 6 4 Notice of Transport and Storage Condi...

Страница 4: ...ge status z User calibration function z Software to connect to PC via USB port z Thermal printer integrated convenient for in field printing z NI MH rechargeable battery as the power source Charge cir...

Страница 5: ...D 15 766 30HLD 15 544 40HLD 15 12 HLD 15 12 HLD 15 5 G 590 40HLG 500 40HLG 12 HLG 12 HLG 6 E 725 30HLE 508 40HLE 12 HLE 12 HLE 7 C 822 30HLC 590 40HLC 12 HLC 12 HLC z Measuring range HLD 170 960 HLD...

Страница 6: ...rage temperature 30 60 Relative humidity 90 The surrounding environment should avoid of vibration strong magnetic field corrosive medium and heavy dust 1 6 Safety Instructions z The instrument can onl...

Страница 7: ...ardness Tester Appearance 1 Main unit 2 Impact device 2 1 2 Parts of the Main body 1 Socket of impact device 2 Socket of USB 3 LOGO 4 Paper compartment cover 5 LCD display 6 Keypad 7 Battery switch 8...

Страница 8: ...t 5 Connection cable 6 Impact body 7 Support ring 2 1 4 Different Types of Impact Device DC DL C D 15 E G 2 2 Main Screen Below is the main display screen Instruction of the Main Display Screen Hardne...

Страница 9: ...Up SC ALE Hardness Scale Selection Direction change Minus or Down FEED Manual Paper Feed Impact Times set Print data Cancel or Exit Data logging or Enter Press key to store present group of measured v...

Страница 10: ...ust be calibrated using a standard hardness block before use as the first time or having not been used for a long time or having reset the instrument system Press key meanwhile pressing down the key t...

Страница 11: ...oth and th o the couple ormed and b y the test back elf should be onform to th ple surface uld be avoid ed surface c ter smoothi ecessary for The sample ce of sampl m D DC device the ss minimum n surf...

Страница 12: ...wo impact po m to the reg from the Le n relations nding hardn e homogene ons which n nding hardne ive curve C the socket o instrument l contact is f ng the impa rmly on the of the impa quired to be l...

Страница 13: ...ing the impact device must be done during Power off Otherwise the main body could not identify the type of the impact device and it may damage the circuit board of the main body z You could not save t...

Страница 14: ...e charge power is connected 5 2 Material Setting Press key to change the material to the one you want to preset Hardness scale recovers to HL automatically after material presetting changed Please sel...

Страница 15: ...o hardness testing is the only selection if the impact device is not D or DC type In hardness testing Press SC ALE key to change the hardness scale The supported hardness scale includes HL HV HB HRC H...

Страница 16: ...an value of the group If there is no data in the memory it will display E04 which means no memory data and then return back 2 Use the key and the key to scroll to the file that will be deleted 3 Press...

Страница 17: ...y operation for 5 minutes it will turn itself off Before powering off the LCD display of the instrument will continue flashing for 20 seconds Except key press any key could stop the twinkle of LCD scr...

Страница 18: ...tery cover z Turn on the instrument to check Warning Please pay much attention to the polarity of the battery during battery replacement 5 14 Connecting to a Computer Insert one connection plug of the...

Страница 19: ...e warranty service can be carried on 6 3 Fault Analysis Evacuation Fault Appearance Fault Analysis Handling method Charge failure Battery failure Replace the battery with a new pack The battery is swi...

Страница 20: ...2 99 5 33 3 99 3 31 8 10 2 1 35 8 102 6 30 6 96 8 Cold work tool steel HRC 20 4 67 1 19 8 68 2 20 7 68 2 22 6 70 2 HV 80 898 80 935 100 94 1 82 1009 Stainless steel HRB 46 5 10 1 7 HB 85 655 HV 85 802...

Страница 21: ...el 500 710 737 1670 3 Cr steel 500 730 707 1829 4 Cr V steel 500 750 704 1980 5 Cr Ni steel 500 750 763 2007 6 Cr Mo steel 500 738 721 1875 7 Cr Ni Mo steel 540 738 844 1933 8 Cr Mn Si steel 500 750 7...

Страница 22: ...5kg 2 5kg 0 05 2kg 5kg 2 5kg 0 05 2kg 1 5kg 0 5 1 5kg 0 02 0 5kg 15kg 5 15kg 0 5 5kg 5kg 2 5kg 0 05 2kg Min thickness of sample Coupling tightly Min layer thickness for surface hardening 5mm 0 8mm 5m...

Страница 23: ...R14 esting cylin surface R2 testing cylin surface R1 testing cylin surface R12 testing cylin surface R16 testing sphe surface SR1 testing sphe urface SR14 testing sphe surface SR1 testing sphe urface...

Отзывы: