background image

 

 

 

©  PCE Instruments 

 

MENU 
ITEM 

DESCRIPTION 

OPTIONS 

FACTORY 
SETTING 

 

Selection of the calibration 
points 

 

USA 
(pH4.01/7.00/10.01) 

USA 

 

NIST 
(pH4.01/6.86/9.18) 

 

Number of calibration points 

 

1 point 

2 points 

 

2 points 

 

3 points 

 

Selection of the temperature 
unit 

 

Celsius 

°C 

 

Fahrenheit 

 

Temperature calibration 

 

--- 

--- 

 

Enables the hold function 
when the measured value is 
stable 

 

On 

Off 

 

Off 

 

The auto-off function switches 
the unit off after 8 minutes 
without any key being pressed 

 

On 

On 

 

Off 

 

Reset to factory settings 

 

Yes 

No 

 

No 

 
After resetting the meter to factory settings, calibration is mandatory. 
 

pH calibration 

The  devices  allow  a calibration  of  up  to  3  points.  To  achieve  high measurement  accuracy,  we 
recommend at least a 2-point calibration. The meter automatically recognises and calibrates to 
the following standard buffer values. 
 
USA standard buffer pH 4.01, 7.00, 10.01 
NIST standard buffer pH 4.01, 6.86, 9.18 
 
The 1-point calibration can only be performed with pH 7.00 or pH 6.86, otherwise the calibration 
will not be accepted. 
The meter must be calibrated before first use.  
Regular calibration is recommended to ensure  measurement accuracy. DO NOT reuse the pH 
buffer solution after calibration as impurities in the solution will affect the calibration and eventually 
the accuracy of the measurement. 
 

 

Содержание PH 26F

Страница 1: ...E PH 28L pH Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last chan...

Страница 2: ...tion 4 5 1 Insert batteries 4 5 2 Preparation of the buffer solution 4 5 3 Electrode control 4 5 4 Switch meter on off 5 5 5 Basic settings 5 6 pH calibration 6 6 1 1 point calibration 7 6 2 2 point c...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...ement Measurement range 0 60 C 32 140 F Accuracy 1 C Resolution 1 C Calibration points 1 General specifications Temperature compensation 0 60 C 32 140 F Temperature of medium 0 60 C 32 140 F Environme...

Страница 5: ...ter PCE PH 28L 1 x buffer 4 01 pH powder for 250 ml solution 1 x buffer 7 00 pH powder for 250 ml solution 1 x buffer 10 01 pH powder for 250 ml solution 3 x container for calibration solution 2 x bat...

Страница 6: ...3 Align the guide groove with the electrode and carefully push the electrode into the meter Screw the union nut of the electrode tight 5 2 Preparation of the buffer solution Prepare 3 sealable contai...

Страница 7: ...key being pressed The function can be enabled or disabled in the basic settings 5 5 Basic settings Press and hold the CAL key for 3 seconds to call up the basic settings Press the CAL key to call up t...

Страница 8: ...ettings calibration is mandatory 6 pH calibration The devices allow a calibration of up to 3 points To achieve high measurement accuracy we recommend at least a 2 point calibration The meter automatic...

Страница 9: ...hows pH 7 00 CAL1 or pH 6 86 CAL1 6 1 3 Immerse the electrode in the pH 7 00 or pH 6 86 buffer solution with the end of the electrode completely immersed in the calibration solution Gently stir the so...

Страница 10: ...electrode in the pH 4 01 or pH 10 01 pH 9 18 buffer solution with the end of the electrode completely immersed in the calibration solution Carefully stir the solution with the electrode to obtain a ho...

Страница 11: ...tely immersed in the calibration solution Stir the solution carefully with the electrode to obtain a homogeneous solution 6 4 4 Press the ENTER key The meter starts the calibration 6 3 5 After the mea...

Страница 12: ...ration 7 Temperature calibration When deviating temperature values are displayed you can adjust them in the corresponding menu item 7 1 Press and hold the CAL key for 3 seconds to call up the basic se...

Страница 13: ...he glass sensitive membrane moist Make sure that the electrode is thoroughly rinsed with distilled or deionised water after use If you do not use the meter for a long period of time store the electrod...

Страница 14: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Страница 15: ...hazlar Ltd ti Unit 11 Southpoint Business Park Via Pesciatina 878 B Interno 6 Halkal Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece...

Отзывы: