Manual
www.industrisal-needs.com
3
2.2
Warning notices
-
This device is only designed for the applications mentioned in this manual. In case the
device is used for any other application than these, it might lead to hazardous situations.
-
Do not expose the device to extreme temperatures, direct sunlight, extreme air humidity
or liquids.
-
The device may be opened only by specialized personnel of PCE instruments
-
Do not operate the device with wet hands.
-
Technical alternations on the device are not allowed.
-
The device should be only cleaned with a damp cloth. Do not use scrubbing agents or
or detergents of any kind.
-
The device may be only combined with the optional equipment provided by PCE
instruments or comparable alternatives.
-
Please check the prior to every operation the enclosure of the device on any visible
damages. In case there appears any visible damage, refrain from any further application
of the device.
-
Furthermore the device may not be applied in any areas where the surrounding
conditions (e.g. temperature, air humidity …) deviate from the specifications and limits
listed in this manual, any application of the device is prohibited.
-
The device cannot be used in explosion-hazardous areas.
-
The threshold specified in this manual cannot be trespassed under any circumstance.
-
Do not use the device, if the battery is down (e.g. the display indicates that the battery is
low), since it might lead to wrong measurement results. First change the batteries
before operating the device again.
-
Please check the measurement device prior to every operation with a standard.
-
Do not place the measurement device anywhere with its control panel facing down (e.g.
control panel facing the table)
-
Do not perform any measurements with the device connected to the computer.
-
Turn off the device before changing the batteries of the device.
-
A disregard of these instructions can lead to a damage of this device and serious
injuries.
For any further information feel free to contact PCE Instruments.
3
Technical Specifications
Measurement range:
-
100 … +400 °C
Accuracy:
± 0.05 % + 0.1 °C (at 18 °C up to 28 °C) only device
Resolution:
0.01 °C
Sensor:
4- terminal Pt-100 class A (cable length approx.. 1.5 m)
Temperature unit:
°C / F
Display:
5
– digit LCD
Data transfer:
USB
Measurement frequency:
0.5 Hz
Power Supply:
9 V bloc battery or optional 9 V adapter
Dimensions:
185 x 65 x 36 mm
Battery lifetime:
approx. 100 hours
Weight:
360 g
Operation conditions:
0° … +50 °C, <80 % rH (non-condensing)
Storing conditions:
-
10 … +60 °C, <80% rH (non-condensing)