6.2.6 

Program sampling menu description 

 

 

 
 
After setting the single sampling quantity, sampling times and sampling interval, press the START 
button  to  start  program  sampling.  The  grouping,  location  and  delay  of  program  sampling  are 
saved by calling up the corresponding parameters of standard sampling directly. The total number 
of samples will be displayed in the data in the next row after setting the values of single sampling 
amount and sampling number. 
After pressing START, the sampling delay is carried out first. After the delay is completed, a single 
sampling  is  started  and  the  sample  size  is  equal  to  the  set  single sampling  amount.  After  the 
sampling  of  single  sampling  quantity  is  completed,  the  sampling  times  will  be  automatically 
reduced by 1 and the sampling will stop. At the same time, the countdown of the sampling interval 
starts. When the countdown of sampling interval changes to 0, a single sampling is started again. 
Do this until the STOP button is pressed or the sampling number becomes 0. 
Setting range of single sampling quantity: 10-3000 (minimum setting amount: 10) 
Setting range of total sampling times: 0-99 
Sampling interval setting range

0-99 

 

Note 1

 

The value of single sampling amount * total sampling times should be less than 6000. If it is higher 
than  6000,  the  set  modification  parameters  will  automatically  change  to  the  corresponding 
minimum setting value. 
 

Note 2

 

The  total  estimated  time  should  be  less  than  99:59:59.  If  it  is  higher  than  this  value,  the  set 
modification parameter will automatically change to the corresponding minimum setting value. 
 
 
 
 
 
 
 
 
 
 
 

 

Содержание PCE-AS1

Страница 1: ...pler PCE AS1 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 16...

Страница 2: ...pplications 2 3 Features 2 4 Specifications 3 4 1 Delivery contents 3 5 Device description 3 6 Operating instructions 5 6 1 Key description 5 6 2 Menu 6 7 Maintenance and precautionary measures 10 8 T...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...ime and sampling quantity of the automatic air sampler can be set directly through the operation panel 6 The sampler can be controlled by Bluetooth function on mobile phone NOTE An App is NOT availabl...

Страница 5: ...dio alarm Yes Procedure cycle sampling Up to 99 continuous sampling times within 100 hours Sampling parameter setting 99 groups and 30 points Recording function 4000 sets of sampling data records Batt...

Страница 6: ...PCE Instruments 4 1 Cover of the collecting head 2 Specimen mount 3 Pump 4 Collecting head stand 5 Control panel with touch screen 6 Power supply 7 On off switch...

Страница 7: ...time is displayed when sampling Bluetooth connection status colour means connected gray means not connected When the buzzer is on no diagonal line means on and a diagonal line indicates off Current b...

Страница 8: ...just the selected setting options When a setting item is selected the key is valid Enter the setting menu of the instrument and set the corresponding parameters in the corresponding interface Start sa...

Страница 9: ...five groups of sampling data You can press the up and down keys to turn the page You can view the first 100 groups of sampling data within the selected time range directly on the instrument at once I...

Страница 10: ...password by using the up and down buttons to the right When the password is correct press the CONFIRM key to enter the flow calibration menu as shown in the second figure When entering the flow calib...

Страница 11: ...antity is completed the sampling times will be automatically reduced by 1 and the sampling will stop At the same time the countdown of the sampling interval starts When the countdown of sampling inter...

Страница 12: ...measures Do not perform sampling with toxic corrosive gas The device is also not intended for use in acidic alkaline environments It is not allowed in areas where there is fine dust exceeding 0 3 mg...

Страница 13: ...dicates that some individual keys do not work The internal computer of the sampler has detected a fault Use the key at the back to turn off the meter Wait 5 seconds until you switch it back on All par...

Страница 14: ...Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 05...

Отзывы: