OPERATING GUIDE
 

To connect the probe 

With the meter turned off, connect the 

HI 62911 

probe to the DIN socket on the top of the meter by 
aligning the pins and pushing in the plug. Tighten the 
nut to ensure a good connection. Remove the 
protective cap from the probe be-fore taking any 
measurements. 

 
To turn the meter ON and check the battery 
status

Press the ON/OFF/MODE button until the display 
lights up. At start-up, all the LCD segments are dis-
played for 1 second, then the percent indication of the 
remaining battery life is shown for another second 
(E.g. % 100 BATT). The meter then enters the normal 
measuring mode. 

Note

:  If the display needs to be checked, keep the 

 

ON button pressed while turning the meter on. 

 

The meter will display all segments as long as 

 

the button is pressed. 

 
To freeze the display 

While in measurement mode, press the SET/HOLD 
button, HOLD appears on the secondary display and 
the reading will be frozen on the LCD (E.g. pH 5.73 
HOLD). Press any button to return to normal mode. 

 
To turn the meter OFF 

While in normal measurement mode, press the 
ON/OFF/MODE button. OFF will appear on the 
secondary display. Release the button. 

 
Note

:  The meter is provided with an acoustic signal 

 

feature which can be disabled using the 

 

switch located in the battery compartment. 

Note

:  When the meter detects the absence of a 

 

temperature probe at its input, the Automatic 

 

Temperature Compensation is turned off, and 

 

the meter uses a default value of 25°C (77°F) 

 

for the temperature measurement and com- 

  

pensation. In this condition, the secondary LCD

 

 

 

shows 25.0°C (77.0°F) blinking. When a probe 

 

is connected, the meter automatically returns 

 

to the ATC mode, the ATC tag is turned on,   

  

and the temperature is shown on the secon- 

  

dary display.

 

Peliminary Examination

 

Remove the instrument from the packing material 
and examine it carefully to make sure that no da-
mage has occurred during shipment. If noticeable 
damage is evident, notify your dealer. 

Note:  

Save all packing material until you

 

are sure that the instrument functions

 

correctly. All defective items must be

 

re-turned in the original packing toge-

  

ther with the supplied accessories. 

 
 
 

General Description 
 
HI 99131 

is a portable, microprocessorbased pH 

and temperature meter with a special probe. The  

HI 62911 

pH electrode has been specially designed 

and fully shielded for applications in plating baths. 
The titanium electrode body acts like a Faraday 
cage, and allows stable readings even in samples 
where strong elec- trical fields are involved. More-
over, a built-in temperature sensor allows simultan-
eous pH and temperature readings and a mini am-
plifier renders measurements impervious to noise 
and electrical interferences. 
 
Each meter comes with: 
• 

HI 62911 

pH/temperature probe 

• 3x1.5V AA alkaline batteries 
• instruction manual 
• plastic carrying case

 

Содержание HI-99131

Страница 1: ...s HI 99131 www warensortiment de PCE Deutschland GmbH Im Langel 4 D 59872 Meschede Deutschland Tel 01805 976 990 Fax 029 03 976 99 29 info warensortiment de www warensortiment de 14 Cent pro Minute aus dem dt Festnetz max 42 Cent pro Minute aus dem dt Mobilfunknetz ...

Страница 2: ...ttery compartment Note When the meter detects the absence of a temperature probe at its input the Automatic Temperature Compensation is turned off and the meter uses a default value of 25 C 77 F for the temperature measurement and com pensation In this condition the secondary LCD shows 25 0 C 77 0 F blinking When a probe is connected the meter automatically returns to the ATC mode the ATC tag is t...

Страница 3: ...rmal mode PH MEASUREMENT CALIBRATION Make sure the meter has been calibrated before use If the probe is dry soak it in HI 70300 storage solution for one hour to reactivate it Place the tip of the probe into the sample to be tested Stir briefly and wait until the symbol on the top left of the LCD is turned off The LCD shows the pH value automatically com pensated for temperature on the primary LCD ...

Страница 4: ... solution 500 mL HI 7007L pH 7 01 buffer solution 500 mL HI 7009L pH 9 18 buffer solution 500 mL HI 7010L pH 10 01 buffer solution 500 mL HI 70300L Storage solution 460 mL HI 7061L Cleaning solution 460 mL HI 721312 Hard carrying case HI 710007 Blue shockproof rubber boot HI 76405 Electrode holder SPEZIFICATIONS Range 2 00 to 16 00 pH 5 0 to 105 0 C 23 0 to 221 0 F Resolution 0 01 pH 0 1 C 0 1 F A...

Отзывы: