3.2.8 

Calibration 

3.2.8.1 

CO

factory calibration 

Touch  the  menu  item  RESET  CO

To  Factory  Settings  to  calibrate  the  CO

sensor  to  factory 

settings. The  meter must  be  placed  in a  well ventilated  room  or  placed  outside  and  run for  15 
minutes. 

3.2.8.2 

Automatic calibration 

The  PCE-CMM  8  has  an  integrated  self-correcting  ABC  algorithm  (Automatic  Baseline 
Correction) for automatic adjustment of the baseline value. This calibration is carried out regularly 
at intervals of 10-15 days. 
For  this,  the  PCE-CMM  8  requires  a  regular  supply  of  fresh  air.  The  measured  value  is 
automatically adjusted. 
The ABC function is permanently switched on (cannot be switched off) and adjusts the baseline 
value at regular intervals when the CO2 level is low. Do not use this meter in a closed environment 
where  the  CO2  concentration  remains  relatively  high.  This  would  negatively  affect  the  self-
correction  of  the  device.  We  therefore  recommend  performing  the  manual  calibration  (3.2.8.1) 
weekly. 

3.2.9 

Offset settings 

3.2.9.1 

CO

offset 

Touch the menu item CO

Sensor Calibration 

 to set an offset for the CO

reading. 

3.2.9.2 

Temperature offset 

Touch the menu item Built In Sensor 

 to set an offset for the temperature reading. 

3.2.9.3 

Humidity sensor offset 

Touch the menu item Humidity Sensor 

 to set an offset for the humidity value reading. 

3.2.10 

Selecting the time zone 

Touch  this  menu  item  to  set  the  time  difference  from  GMT.  Press  the 

 key  to  enable 

synchronisation  of  the  power  on/off  time  with  the  internet  time.  To  do  this,  the  meter  must  be 

successfully connected to the internet. 

3.2.11 

Daylight saving time 

Touch the menu item 

 to set the daylight saving time based on the location. 

3.2.12 

Home screen colour 

In the menu item Home Screen Colour 

, select whether the background colour should be blue 

or black. 
 

 

Содержание CMM 8

Страница 1: ...CO2 Analyser User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change 25...

Страница 2: ...description 2 2 1 Technical specifications 2 2 2 Display touch screen 3 2 3 Dimensions 3 3 Operating instructions 4 3 1 On off 4 3 2 Settings 4 3 3 Measurement graph data history 6 3 4 Alarms 6 3 5 CO...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...Resolution 10 50 C 14 122 F 1 C 0 1 C Measured variable air humidity Measurement range Accuracy Resolution 20 95 RH 5 RH 1 RH General technical specifications Display 3 5 touch LCD Alarm display Traf...

Страница 5: ...lower or upper alarm 3 CO2 reading changes ist colour 4 Increase or decrease alarm value 5 Lock screen 6 Unit change C F 7 Air humidity reading in RH 8 Show progress diagramme 9 Temperature reading 10...

Страница 6: ...ature Format to switch between C and F 3 2 4 Setting the display brightness Touch the menu item Backlight to adjust the display brightness using the arrow keys at the bottom right of the display If yo...

Страница 7: ...ere the CO2 concentration remains relatively high This would negatively affect the self correction of the device We therefore recommend performing the manual calibration 3 2 8 1 weekly 3 2 9 Offset se...

Страница 8: ...Read data history on the PC Connect the meter to a PC using the USB cable The CO2 meter is recognised as a mass storage device In the file DATALOG TXT you can view the previously saved measurement dat...

Страница 9: ...ww pce instruments com english terms 5 Disposal For the disposal of batteries in the EU the 2006 66 EC directive of the European Parliament applies Due to the contained pollutants batteries must not b...

Страница 10: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Отзывы: