4.  Close the head of the measurement chamber again and make sure that it is not too tight.  
5.  Compress the sample within the sawdust humidity detector by using the compacting device.  

 

 

After operating the compacting device a couple of times, an indicator will show up to give 
information on the compression pressure. The appropriate compression is reached when a red 
ring around the indicator is shown. The red ring should only point out of the head of the 
measuring chamber by a maximum of 0.5 to 1 millimetre. 

 

 
 
 
 
 
Red ring 

 

Please be advised that the device may be damaged if the compression pressure is too high!  

 

4.3  Make a measurement 

1.  Press and hold the ON/OFF button of the device to start it.  
2.  Press the MATERIAL button and adjust your desired sample material. 

 

Material 

Adjustment 

Sawdust, wood chips 

Straw, hay 

 

3.  Press the Temperature button to set the temperature the sample material has. Each time you 

press the button, the temperature is increased in steps of 2 °C. By pressing and holding the 
button, the speed of the increase will be higher. A difference between the set temperature and 
the actual temperature of ±4 °C will not affect the measurement result.  

 
 

 

Содержание PCE-WT 1N

Страница 1: ...piter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Busin...

Страница 2: ...Safety notes 3 3 Specification 3 4 System description 4 4 1 Theoretical determination of humidity 4 4 2 Preparations before measurements 4 4 3 Make a measurement 5 4 4 General information 7 5 Disposa...

Страница 3: ...rying case Instruction manual 2 Safety notes Please read this manual carefully and completely before you use the device for the first time The device may only be used by qualified personnel and repair...

Страница 4: ...surements 1 Open the head of the measurement chamber 2 Pull back the compacting device completely in order to ensure that the measurement chamber can take the maximum volume 3 Fill the measurement cha...

Страница 5: ...maximum of 0 5 to 1 millimetre Red ring Please be advised that the device may be damaged if the compression pressure is too high 4 3 Make a measurement 1 Press and hold the ON OFF button of the devic...

Страница 6: ...measurement may take up to 10 seconds Moist samples will give a result within one or two seconds 5 Pull back the press to its completely again and open the head 6 Use the compacting device again to pu...

Страница 7: ...atability The size of the sample has effects on the measurement results The smaller the sample size the more precise the measurement result pellets can be crushed with a hammer You can also increase t...

Страница 8: ...use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 6 Contact If you have any questions about our rang...

Отзывы: