PCE Americas PCE-318 Скачать руководство пользователя страница 7

. CALIBRATION PROCEDURES

Acoustic Calibration with sound calibrator TES-1356

Ov

Un

Un

Ov

1. Turn off the sound calibrator.

2. Press 

 key turn on the meter.

3. Insert the microphone very carefully and slowly all the way

into the sound calibrator coupler.

4. Set the power switch of the sound calibrator to 94dB.

5. Adjust the CAL potentiometer of the meter, until display

reading is 94.0dB.

6. Set the power switch of the sound calibrator to OFF.

7. Remove the microphone very carefully and slowly from the

coupler.

-10-

. MEASUREMENT

10-1 When using this meter, all processing functions provided

by  the  meter  are  carried  out  simultaneously.    For
example, when sound level measurement is selected, the
maximum sound level are also determined.

Sound level measurement :

1.  Press 

 key turn on the meter, the default settings is

A frequency weighting, fast time weighting and 60 to
120dB sound level range.

2.  Press “A” or “C” keys to select desired frequency

weighting.    For normal sound level measurements,
select “A” setting.

3.  Press  “FAST”  or  “SLOW”  keys  to  select  desired  time

weighting  (dynamic  characteristics).    Normally  the
“FAST” setting should be used.

4.  When performing  measurements  according  to  IEC  or

other  standards,  the  frequency  and  time  weighting
setting required by the standard should be selected.

5.  Press  “

”  or  “

 

 

”  keys  to  select  the  desired  level

range.    Choose  a  setting  in  which  the  bar  graph
indication  to  about  the  middle  of  the  range.    If  the
“ 

OV

 ” (Over) or “ 

UN 

and 

 

  

 

 

  

 

” (Under) indicates

light up frequently, change the level range setting.

6.  The numeric level indication shows the currently

measured sound level and the elapsed measurement
time.    The reading is update once every second.

7.  Press “MAX” key to cycle switch the currently sound

level with elapsed measurement time 

 the maximum

sound level (“max” is shown) with measured elapsed
time 

 the minimum sound level ( “min” is shown) with

measured elapsed time.

-11-

Содержание PCE-318

Страница 1: ...1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info industrial needs com ...

Страница 2: ...not covered in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Do not use abrasives or solvents on this instruments Safety symbols Comply with EMC U S Pat No Des 483 680 Ⅱ Ⅱ Ⅱ Ⅱ PRECAUTIONS Do not try to remove the microphone grid you can easily damage the microphone in this way Protect the meter from impact Do not drop it Transport it in the...

Страница 3: ...in with elapsed time Both AC and DC signal output is available from a single standard 3 5mm coaxial socket suitable for a frequency analyzer level recorder FFT analyzer graphic recorder etc 2 Ⅳ Ⅳ Ⅳ Ⅳ PARAMETERS The following parameters are used on the meter LA A frequency weighting sound pressure level LC C frequency weighting sound pressure level FAST Fast time weighting SLOW Slow time weighting ...

Страница 4: ...resolution Bar graph indication of current sound level 1dB resolution Elapsed time display 100 hours maximum Memory and read address display 99 data sets maximum 4 Warning indications Over range indication OV displayed at upper limit of the range Un displayed at lower limit of the range Outputs AC output using selected frequency weighting Output voltage 1Vrms at full scale of the range Output impe...

Страница 5: ... key Press to read the stored data sets in memory press again to exit read mode 9 keys Level range keys select the level range for the measurement The following six settings are available 20 to 80 30 to 90 40 to 100 50 to 110 60 to 120 70 to 130 When read mode the keys select the memory address to be displayed 10 DC AC output AC signal with frequency weighting DC signal corresponding to sound leve...

Страница 6: ...y and before connecting the AC adaptor be sure to turn off the meter 1 Battery Loading Before replace the batteries must press key turn off the meter Remove the cover of the battery compartment Insert the new 9V battery Replace the battery cover Press key turn on the meter then press key turn off the meter 2 Battery capacity indicator When use battery operating the meter periodically check this in...

Страница 7: ... fast time weighting and 60 to 120dB sound level range 2 Press A or C keys to select desired frequency weighting For normal sound level measurements select A setting 3 Press FAST or SLOW keys to select desired time weighting dynamic characteristics Normally the FAST setting should be used 4 When performing measurements according to IEC or other standards the frequency and time weighting setting re...

Страница 8: ...h the logged readings Press MAX key to cycle switch the stored currently sound level with elapsed measurement time the maximum sound level max is shown with measured elapsed time the minimum sound level min is shown with measured elapsed time Press READ key again to exit READ mode 12 3 To clean the memory Press key to turn off the meter Press and hold down MEM key then press key to turn on the met...

Отзывы: