PCB Load&Torque 1380-01A Скачать руководство пользователя страница 6

 

PCB Load & Torque, Inc.   

 Toll-Free in USA 866-684-7107   

 716-684-0001   

 www.pcbloadtorque.com

 

 

 

ROD END LOAD CELL OPERATION MANUAL 

 

1

 

TABLE OF CONTENTS 

 

1.0 INTRODUCTION.................................................................................................................. 2

 

 

2.0 SAFETY PRECAUTIONS .................................................................................................... 2

 

 

3.0 OVERVIEW .......................................................................................................................... 2

 

3.1 Dimensions .......................................................................................................................................2

 

3.2 Optional Components .......................................................................................................................2

 

 

4.0 MECHANICAL INSTALLATION .......................................................................................... 2

 

 

5.0 ELECTRICAL INSTALLATION ........................................................................................... 3

 

5.1 Electrical Drawing / Western Regional Std. ......................................................................................3

 

5.2 Output Polarity ..................................................................................................................................3

 

5.3 Cable & Grounding Considerations ..................................................................................................3

 

 

6.0 CALIBRATION..................................................................................................................... 3

 

6.1 Calibration Certificate Description.....................................................................................................3

 

6.1.1 Measured Output.....................................................................................................................4

 

6.1.2 Hysteresis................................................................................................................................4

 

6.1.3 Best Fit Output ........................................................................................................................4

 

6.1.4 Strain Gage Measurements .....................................................................................................4

 

6.1.5 Shunt Calibration Standard Resistor .......................................................................................4

 

6.1.6 Static Error Band (SEB)..........................................................................................................4

 

 

7.0 SHUNT CALIBRATION DESCRIPTION.............................................................................. 4

 

7.1 Resistor Value ...................................................................................................................................4

 

7.2 Shunt Calibration Process.................................................................................................................4

 

7.3 Estimating Shunt Resistor for a Given Load .....................................................................................4

 

 

8.0 MAINTENANCE................................................................................................................... 5

 

 

9.0 TROUBLE SHOOTING ........................................................................................................ 5

 

9.1 Mechanical Trouble Shooting ...........................................................................................................5

 

9.2 Electrical Trouble Shooting ...............................................................................................................5

 

 

10.0 CALIBRATION / REPAIR SERVICES ............................................................................... 5

 

10.1 RMA / Purchase Order....................................................................................................................5

 

 

11.0 WARRANTY ...................................................................................................................... 5

 

 
 
 
 
 
 
 
 

Содержание Load&Torque 1380-01A

Страница 1: ...allation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbloa...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...ounding Considerations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 4 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Cal...

Страница 7: ...od end load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing fixtures be sure to thread t...

Страница 8: ...The instrument or control system is grounded through its power cord Ground loops and measuring system wiring may result in unstable or noisy signals A simple test with a voltmeter connected between th...

Страница 9: ...load on a load cell The calibration certificate will indicate which leg of the bridge to apply the shunt resistor to for both tension and compression load simulation Typically tension is simulated by...

Страница 10: ...ignal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 10 0 CALIBRATION REPAIR SERVICES PCB Load Torque offers calibration and...

Страница 11: ...ing Thread 3 4 16 Female No Metric Equivalent Cover Material Anodized Aluminum Anodized Aluminum Sensor Material Anodized Aluminum Anodized Aluminum Sensing Element Strain Gage Strain Gage Deflection...

Страница 12: ...10 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 45488 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to o...

Отзывы: