background image

Model Number

M1631-06C

STRAIN GAGE LOAD CELL

Revision: E
ECN #: 43825

Performance

ENGLISH

SI 

Measurement Range

4496 lb

20 kN

[5]

Sensitivity(± 10 %)

2 mV/V

2 mV/V

[6]

Non-Linearity

≤ 0.15 % FS

≤ 0.15 % FS

Hysteresis

≤ 0.15 % FS

≤ 0.15 % FS

Non-Repeatability

≤ 0.05 %RO

≤ 0.05 %RO

Resonant Frequency

2.5 kHz

2.5 kHz

Creep(in 20 minutes)

≤ 0.08 %

≤ 0.08 %

Environmental

Overload Limit

6745 lb

30 kN

Load Limit(Side Force, F

X

 or F

Y

)

450 lb

2000 N

[3]

Load Limit

(Bending Moment, M

X

 or M

Y

)

1124 in-lb

127 Nm

[3]

Load Limit

(Axial Torque, M

Z

)

1124 in-lb

127 Nm

[3]

Temperature Range(Operating)

0 to +200 °F

-18 to +93 °C

Temperature Range

(Compensated)

+75 to +150 °F

+21 to +65 °C

Temperature Effect on Output(Maximum)

± 0.002 %Reading/°F

± 0.0036 %Reading/°C

[4]

Temperature Effect on Zero Balance(Maximum)

± 0.0015 %FS/°F

± 0.027 %FS/°C

[4]

Electrical

Bridge Resistance

350 Ohm

350 Ohm

[1]

Excitation Voltage(Recommended)

10 VDC

10 VDC

[2]

Insulation Resistance

>5x10

9

 Ohm

>5x10

9

 Ohm

Zero Balance

≤ 1 % FS

≤ 1 % FS

Output Polarity

+Tension

+Tension

Physical

Size (Length x Height x Width)

2.50 in x 3.50 in x 1.50 in

64 mm x 89 mm x 38 mm [7]

Weight

2.85 lb

1.3 g

Housing Material

Nickel Plated Steel

Nickel Plated Steel

Sensing Element

Strain Gage

Strain Gage

Deflection at Full Scale Capacity

0.003 in

0.05 mm

Electrical Connector

10 ft - Integrated Cable

10 ft - Integrated Cable

Electrical Connection Position

Side

Side

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

NOTES:

[1]Nominal.
[2]Calibrated at 10 VDC, usable 5 to 20 VDC or VAC RMS.
[3]Singularly applied, i.e. no other extraneous loads.
[4]Over compensated operating temperature range.
[5]FS - Full Scale.
[6]RO - Rated Output.
[7]See Outline Drawing 46141 for Complete Dimensions

PCB Load & Torque, Inc.
24350 Indoplex Circle
Farmington Hills, MI 48335
UNITED STATES
Phone: 866-684-7107
Fax: 716-684-0987
E-Mail: [email protected]
Web site: 
http://www.pcbloadtorque.com

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard 

model except where noted below. More than one option may be used.

Entered: AP

Engineer: PE

Sales: JC

Approved: JSD

Spec Number:

Date: 2/16/2015

Date: 2/16/2015

Date: 2/16/2015

Date: 2/16/2015

46017 

Содержание Load & Torque M1631-06C

Страница 1: ...stallation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbl...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...erations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 3 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Stand...

Страница 7: ...STALLATION Loads applied to S type load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing...

Страница 8: ...or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded I...

Страница 9: ...ision shunt resistor is 60 000 ohms 0 1 simulates an output of approximately 73 of the full scale output for the load cell The calibration values for each bridge are found on the calibration certifica...

Страница 10: ...aboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point descending Additional data points are avail...

Страница 11: ...x 89 mm x 38 mm 7 Weight 2 85 lb 1 3 g Housing Material Nickel Plated Steel Nickel Plated Steel Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 003 in 0 05 mm Electrical C...

Страница 12: ...ODEL M1631 06C S CELL JAM 6 29 11 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 46142 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction...

Отзывы: