Drawing Number: 21075 
Revision: C 
ECN Number: 47023 

 

OPERATION MANUAL FOR 

QUARTZ PRESSURE SENSORS 

SERIES 102 
SERIES 113 

threaded  adaptor  with  floating  clamp  nut  to  allow 
adjustment  of  diaphragm  mounting  depth  where  it  is 
necessary to adapt to various wall thicknesses.  These 
models  are  supplied  only  as  low-pressure  (250  psi 
and 100 psi) sensors and are also "off ground". 
 
Models  102A21  and  102A22  are  high-temperature 
ICP

® 

versions  to  400 

F  (204 

C),  with  a  3/8-24 

straight  threads  adaptor  and  1/8-27  NPT  adaptor, 
respectively. 

 

Series  102:    Thread  Mount  Design,  Ground-
Isolated Sensor 

 

3.0  INSTALLATION 

 
This  manual  contains  outline  and  installation 
information for your specific model. 
 
Prepare 

mounting 

ports 

in  accordance  with 

instructions  given  in  specific  installation  drawings, 
paying particular attention to sealing surfaces.  These  
surfaces must be smooth and free from chatter marks, 
nicks  and  other  irregularities  which  could  preclude a 
pressure tight seal. 
 
To  fully  realize  the  high-frequency  response 
capabilities  of  this  sensor  series,  flush  mounting  of 
the diaphragm must be used. 
 
In some cases, where flash temperatures such as those 
generated  by  blasts  and  shock  fronts  are  present,  it 
may be necessary to thermally insulate the diaphragm 
to minimize signals generated by these effects. 
 

Common black vinyl electrical tape has been found to 
be  an  effective  insulating  material  in  many  cases.  
One  or  more  layers  may  be  used  across  the  end  of 
diaphragm and adaptor. 
 
A  silicone  rubber  coating  approximately  .010”  thick 
has  also  been  proven  effective  in  many  applications.  
General  Electric  RTV  type  106  is  recommended.  
Apply  the  rubber  coating  to  the  surface  of  the 
diaphragm and allow it to cure in accordance with the 
manufacturer’s instructions. (If you have ordered the 
CA option,

 

ablative coated models, further protection 

will not be necessary.) 
 
Although  ICP

®

  sensors  have  low-output  impedance 

and  in  general  are  not  affected  by  moisture,  in 
extreme  environments  it  is  good  practice  to  protect 
cable connections with shrink tubing. 
 
It  is  not  necessary  to  use  low-noise  cable  with  this 
sensor  series.    In  fact,  an  optional  Model  070B09

 

Solder Connector Adaptor allows the use of ordinary 
two-wire cable if desired. 

 

4.0 

OPERATION

 

 
It is only necessary to supply the sensor with a 2 to 20 
mA  constant  current  at  +20  to  +30  VDC  through  a 
current-regulating  diode  or  equivalent  circuit.  (See 
guide  G-0001B  for  powering  and  signal  utilization 
information pertaining to all ICP

®

 instrumentation). 

 
Most of the signal conditioners manufactured by PCB 
have  an  adjustable  current  feature  allowing  a  choice 
of  input  currents  from  2  to  20  mA.    In  general,  for 
lowest  noise  (best  resolution),  choose  the  lower 
current  ranges.    For  driving  long  cables  (to  several 
thousand  feet),  use  higher  current,  up  to  20  mA 
maximum. 
 
To operate system using a PCB signal conditioner: 
 
1. 

Switch power on. 

 
2. 

Wait several minutes for the IC ampli- 

fier to turn on and stabilize. 
 
3. 

Proceed with measurements. 

 

Содержание M102B18

Страница 1: ...Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail i...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...shows the components of the basic ICP probe i e the piezoelectric element and the ICP source follower amplifier These components are joined together as an inseparable sealed assembly at the factory Di...

Страница 7: ...al in many cases One or more layers may be used across the end of diaphragm and adaptor A silicone rubber coating approximately 010 thick has also been proven effective in many applications General El...

Страница 8: ...are also features of these sensors As mentioned previously the diaphragm must be flush mounted to fully realize the high frequency response capabilities of this series 8 0 CALIBRATION Piezoelectric se...

Страница 9: ...elded Hermetic Welded Hermetic Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 50 oz 14 3 gm All specifications are at room temperature unless otherwise specified In the interest o...

Страница 10: ......

Отзывы: