background image

 

 
For these reasons, it is recommended that a recalibration 
cycle  be  established  for  each  accelerometer.    This 
schedule  is  unique  and  is  based  on  a  variety  of  factors, 
such  as  extent  of  use,  environmental  conditions, 
accuracy requirements, trend information obtained from 
previous  calibration  records,  contractual  regulations, 
frequency of “cross-checking” against other equipment, 
manufacturer  recommendation,  and  any  risk  associated 
with incorrect readings.  International standards, such as 
ISO 10012-1, provide insight and suggested methods for 
determining  recalibration  intervals  for  most  measuring 
equipment.    With  the  above  information  in  mind  and 
under  “normal”  circumstances,  PCB  conservatively 
suggests  a 12- to 24-month recalibration cycle for most 
piezoelectric accelerometers.   
 

Note:

 

It  is  good  measurement  practice  to  verify  the 

performance  of  each  accelerometer  with  a  Handheld 
Shaker or other calibration device before and after each 
measurement.    The  PCB  Model  394C06  Handheld 
Shaker  operates  at  a  fixed  frequency  and  known 
amplitude  (1.0  g)  to  provide  a  quick  check  of  sensor 
sensitivity.

 

 

7.1 SENSOR RECALIBRATION 

Accelerometer  recalibration  services  are  typically 
performed  by  PCB’s  internal  metrology  laboratory.  
(Other  international  and  private  laboratories  are  also 
available.)  The PCB laboratory is certified to ISO 9001, 
accredited  by  A2LA  to  ISO  17025,  complies  with  ISO 
10012-1  (and  former  MIL-STD-45662A),  and  uses 
equipment directly traceable to N.I.S.T.  This assures an 
accurate calibration of relevant specifications. 
 
In  addition,  many  companies  choose  to  purchase  the 
equipment  necessary  to  perform  the  recalibration 
procedure  themselves.    While  this  may  result  in  both  a 
savings of time and money, it has also been attributed to 
incorrect  readings  and  costly  errors.    Therefore,  in  an 
effort  to  prevent  the  common  mistakes  associated  with 
customer-performed  calibration,  this  document  includes 
a  broad  overview  of  the  Back-to-Back  Calibration 
technique.    This  technique  provides  a  quick  and  easy 
method  for  determining  the  sensitivity  of  a  test 
accelerometer over a wide frequency range.   

 
7.2 BACK-TO-BACK CALIBRATION THEORY 

Back-to-Back  Calibration  is  perhaps  the  most  common 
method  for  determining  the  sensitivity  of  piezoelectric 
accelerometers.    This  method  relies  on  a  simple 
comparison  to  a  previously  calibrated  accelerometer, 
typically referred to as a reference standard.   
 
 

 

 

 

Figure 9.

 Reference Standard Accelerometer 

 
These  high-accuracy  devices,  which  are  directly 
traceable  to  a  recognized  standards  laboratory,  are 
designed  for  stability,  as  well  as  configured to accept a 
test accelerometer.  By mounting a test accelerometer to 
the  reference  standard  and  then  connecting  this 
combination to a suitable vibration source, it is possible 
to  vibrate  both  devices  and  compare  the  data  as  shown 
in Figure 10.  (Test set-ups may be automated and vary, 
depending  on  the  type  and  number  of  accelerometers 
being calibrated.) 
 

 

 
Figure 10.

 Typical Back-to-Back Calibration System 

 
Because  the  acceleration  is  the  same  on  both  sensors, 
the ratio of their outputs (V

T

/V

R

) must also be the ratio 

of  their  sensitivities.    With  the  sensitivity  of  the 
reference  standard  (S

R

)  known,  the  exact  sensitivity  of 

the  test  sensor  (S

T

)  is  easily  calculated  by  using  the 

following equation: 
 
 

 

S

= S

R

 (V

T

/V

R

 
By  varying  the  frequency  of  the  vibration,  the  sensor 
may  be  calibrated  over  its  entire  operating  frequency 
range.    The  typical  response  of  an  unfiltered 
accelerometer is shown in Figure 11. 

Содержание J357B01

Страница 1: ... accel 15 pC g 9k Hz 10 32 side Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ity regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme shock load or other environmental influence or prior to any critical test PCB Piezotronics maintains an ISO 9001 certified metrology laboratory and offers calibration services which are accredited by A...

Страница 3: ...urchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers should direct all inquiries to their local distributor or sales office A complete list of distributors and offices can be found at www pcb com Customers within the United States may contact their local s...

Страница 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Страница 6: ... F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high temperatures preclude the use of accelerometers with built in microelectronics Enclosed is a Specification Sheet which lists the complete performance characteristics of the particular accelerometer 3 0 ...

Страница 7: ...surface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the mounting surface and spread on a light film of grease oil or similar coupling fluid prior to installation Figure 3 Mounting Surface Lubrication Adding a coupling fluid improves vibration transmissibility...

Страница 8: ...for most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are questionable 5 0 CABLING Care and attention to installation is essential as the reliability and accuracy of your system is no better than that of the output cable Charge mode sensors require the use of ...

Страница 9: ...to a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitivity in line charge converters such as the PCB Series 422 the system sensitivity mV g is determined as the product of the charge amplifier sensitivity mV pC and the charge sensitivity of the accelerom...

Страница 10: ...edure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes associated with customer performed calibration this document includes a broad overview of the Back to Back Calibration technique This technique provides a quick and easy method for determining the sensit...

Страница 11: ...2 Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the test sensor to a reference standard designed with a smooth flat mounting surface In addition to mounting the selection of the proper equipment is critical Some of the more important considerations inclu...

Страница 12: ...rs Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz or 2 kHz Since large sensors may affect high frequency accuracy verify that the test sensor does not mass load the reference standard Validate your calibration system with another accelerometer prior...

Страница 13: ...he interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as listed for the standard model except where noted below More than one option may be used J Ground Isolated Frequency Range 5 7 kHz 7 kHz Frequency Range 10 8 kHz 8 ...

Страница 14: ...IONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M8 X 1 25 MOUNTING INSTRUCTIONS ENGLISH DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 3 7 NEWTON METERS 2 5 FT POUNDS M8 X 1 25 MOUNTING HOLE PREPARATION 6 75 266 X 8 64 340 MIN M8 X 1 25 6H X ...

Страница 15: ......

Отзывы: