Model Number

RH201A 76

ICP® FORCE SENSOR

Revision: B
ECN #: 52977

P erfo rm an ce

EN G L ISH

SI

Sensitivity(± 15 %)

1 mV/lb

224.8 mV/kN

Measurement Range(Compression)

5,000 lb

22.24 kN

Maximum Static Force(Compression)

5,000 lb

22.24 kN

Broadband Resolution(1 to 10,000 Hz)

0.10 lb-rms

0.445 N-rms

[1]

Low Frequency Response(- 5 %)

0.0003 Hz

0.0003 Hz

[2]

Upper Frequency Limit

90 kHz

90 kHz

[3]

Non-Linearity

≤ 1 % FS

≤ 1 % FS

[4]

En viro n m en tal

Temperature Range

-65 to +250 °F

-54 to +121 °C

Temperature Coefficient of Sensitivity

≤ 0.03 %/°F

≤ 0.054 %/°C

Electrical

Discharge Time Constant

≥ 2,000 sec

≥ 2,000 sec

Excitation Voltage

20 to 30 VDC

20 to 30 VDC

Constant Current Excitation

2 to 20 mA

2 to 20 mA

Output Impedance

≤ 100 Ohm

≤ 100 Ohm

Output Bias Voltage

8 to 14 VDC

8 to 14 VDC

Spectral Noise(1 Hz)

0.00307 lb/√Hz

0.0137 N/√Hz

[1]

Spectral Noise

(10 Hz)

0.00115 lb/√Hz

0.00513 N√Hz

[1]

Spectral Noise

(100 Hz)

0.000402 lb/√Hz

0.00179 N/√Hz

[1]

Spectral Noise

(1,000 Hz)

0.000114 lb/√Hz

0.000509 N/√Hz

[1]

Output Polarity(Compression)

Positive

Positive

P h ysical

Preload

1,000 lb

4.448 kN

Stiffness

16 lb/µin

2.8 kN/µm

[1]

Size (Diameter x Bolt Diameter x Sensing
Surface x Height)

0.750 in x 0.25 in x 0.500 in x

0.200 in

19.05 mm x 6 mm x 12.7 mm x

5.08 mm

Size - ID(Hole Diameter)

0.251 in

6.38 mm

Size - OD(Sensor)

0.750 in

19.05 mm

Weight

0.25 oz

7 gm

Housing Material

Stainless Steel

Stainless Steel

Sealing

Hermetic

Hermetic

Electrical Connector

5-44 Coaxial Jack

5-44 Coaxial Jack

Electrical Connection Position

Side

Side

[5]

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

 is a registered trademark of PCB Piezotronics, Inc.

This model, designated with an RH prefix, is RoHS compliant. For further details, and to obtain PCB's RoHS Statement of

Conformance, please visit http://www.pcb.com

M

 - Metric Mount

Supplied Accessory: Model M081A11 Mounting Stud, M5 x 0.8 x 0.730, BeCu (1) replaces Model

081A11

Supplied Accessory: Model M083A15 Pilot Bushing, M5 x 0.8 (1) replaces Model 083A15

N

 - Negative Output Polarity

Output Polarity(Compression)

Negative

Negative

W

 - Water Resistant Cable

Electrical Connector

Sealed Cable

Sealed Cable

Electrical Connection Position

Side

Side

OP TION A L  VERSION S

Optional versions have identical specifications and accessories as listed for the standard model except

where noted below. More than one option may be used.

N OTES:

[1]Typical.
[2]Calculated from discharge time constant.
[3]Estimated using rigid body dynamics calculations.
[4]Zero-based, least-squares, straight line method.
[5]See PCB Declaration of Conformance PS023 for details.

Date: 07/08/2022

Date: 07/08/2022

Date: 07/08/2022

Date: 07/08/2022

37137

Entered: ND

Engineer: MJK

Sales: KWW

Approved: AJA

Spec Number:

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

3425 Walden Avenue, Depew, NY 14043

SU P P L IED A C C ESSORIES:

Model 081A11 Mounting Stud, 10-32 x 0.730, BeCu (1)
Model 083A15 Pilot Bushing, 10-32 (1)

Содержание ICP RH201A76

Страница 1: ...ensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail in...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...tallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description o...

Страница 7: ...ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Inst...

Страница 8: ...simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the spe...

Страница 9: ...eturning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for com...

Страница 10: ...t facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surface...

Страница 11: ...he left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total for...

Страница 12: ...itor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accom...

Страница 13: ...hundreds of feet meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent...

Страница 14: ...ant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents...

Страница 15: ...urned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specificat...

Страница 16: ...cal Connector 5 44 Coaxial Jack 5 44 Coaxial Jack Electrical Connection Position Side Side 5 All specificationsare at room temperature unlessotherwise specified In the interest of constant product imp...

Страница 17: ...17 1 1 2 2 A A B B 16 3 9 187 4 75 170 4 32 200 001 5 08 03 75 19 0 50 12 7 1 48 37 5 73 18 5 MODEL 081A11 STUD 10 32 UNF 2A TYP BOTH ENDS SUPPLIED MODEL 083A15 PILOT BUSHING SUPPLIED MOUNTING HOLE P...

Отзывы: