1.  INTRODUCTION AND DESCRIPTION 
 

The Model 401B04 is a unity gain, non-inverting, impedance-converting voltage amplifier with the same 
characteristics as the built-in ICP

®

 amplifiers provided in most PCB pressure sensors, force sensors, and 

accelerometers.  This model consists of MOSFET circuitry packaged in a convenient Pomona box with BNC 
connectors.  When used in conjunction with a signal generator or oscillator, it can be used in place of an ICP 
sensor to calibrate ICP

®

 sensor signal conditioners, do system verifications and gain adjustments, or check 

frequency response of systems with long input cables.  (The Model 401B04 simulates the circuit of an ICP

®

 

sensor while the signal generator takes the place of the sensor’s crystals.) 
 
 

2.  INSTALLATION AND OPERATION 
 

CAUTION:  Connect the output only to an ICP

®

 type, current-limited signal conditioner. 

 
WARNING: Excessive current may damage this model.  Measure the ICP

®

 sensor signal conditioner 

current prior to connecting the 401B04.  The ICP

®

 excitation current should be 

20 mA. 

 

2.1  CALIBRATION OF ICP

®

 SENSOR SIGNAL CONDITIONERS 

 

Refer to the acceptance test procedure for your particular PCB signal conditioner, dual mode charge 
amplifier, or digital peak meter model.  (Acceptance test procedures are available for most PCB electronic 
products.  Consult factory for details.) 
 
Connect a signal generator or oscillator to the input jack of the 401B04, after observing the caution and 
warning above.  Connect the output jack of the 401B04 to the ICP

®

 sensor signal conditioner being 

tested.  (Various cables and adaptors are available from PCB as needed.)  Switch on the signal 
conditioner and note the output bias voltage on the meter as you would for any ICP

®

 sensor.  The meter 

should read mid scale, indicating that the 401B04 is operational.  Follow the acceptance test procedure 
for your particular PCB model.  The exact gain factor (near unity) of the 401B04 is noted on the 
calibration certificate supplied with the unit.  The nominal value of Gain = 1.0 is on the housing of the unit. 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

401B04 Simulator 

Frequency Counter 

Function Generator 

Digital Volt Meter 

Oscilloscope 

Signal Conditioner 

(Unit Under Test) 

401B04 Simulator 

Frequency Counter 

Function Generator 

Digital Volt Meter 

Oscilloscope 

Signal Conditioner 

(Unit Under Test) 

401B04 Simulator 

Frequency Counter 

Function Generator 

Digital Volt Meter 

Oscilloscope 

Signal Conditioner 

(Unit Under Test) 

Отзывы: