Drawing Number: 21102 
Revision: A 

 

 

 

 

OPERATION MANUAL FOR 

3  

Series 138 

 

 

The  steel  wire  can  have  a  weight  attached  to  it  to 
submerge  the  entire  assembly.  Tape  or  attach  the 
coaxial  cable  loosely  to  the  steel  wire  at  ~one  foot 
intervals. Attach a monofilament line (~50 lb. test) to 
the end of a steel wire braid, and let it hang down past 
the end of the sensor. 

 

Attach  a  three  to  five  lb.  weight  to  the  end  of  the 
monofilament 

and 

tape 

the 

sensor 

to 

the 

monofilament. 

 

Another method of mounting the sensor is to tape the 
sensor  to  the  surface  of  a  structure.  (The  pressure 
measured when mounted to the surface of a structure 
will be twice the free field pressure since the pressure 
wave  reflected  off  the  surface  adds  to  the  pressure 
wave initially striking the sensor.) 

 

For applications in which the 138 is immersed only a 
few  feet  into  a  liquid,  and  the  blast  may  be  in  very 
close  proximity  to  the  sensor,  a  steel  pipe  can  be 
placed  around  the  cable  and  sensor  so  that  only  the 
tourmaline  sensing  element  is  exposed  to  the  blast. 
Since the static pressure is not high, a standard micro 
cable  with  10-32  coax  connector  can  be  used  with 
heat shrink tubing to seal the cable/connector. 

 

3.0 

OPERATION 

 

It is necessary only to supply the sensor with a 2 to 20 
mA  constant  current  at  +20  to  +30  VDC  through  a 
current-regulating diode or equivalent circuit.  

 

Most  PCB  signal  conditioners  have  an  adjustable 
current  feature  allowing  a  choice  of  input  currents 
from 2 to 20 mA. 

 

In  general,  for  best  resolution,  choose  the  lower 
current ranges, and for driving long cables (to several 
thousand  feet)  use  the  higher  current,  up  to  20  mA 
maximum.  Line-impedance  matching  may  be 
necessary  to  produce  flat-frequency  response  over 
long (hundreds or thousands of feet) cables. 

 

Switch  power  on  and  observe  reading  of  bias 
monitoring  voltmeter  on  front  panel  of  the  signal 
conditioner. 

If indicator is in green section of indicator  panel, the 
IC  amplifier  is  producing  proper  bias  (+8  to  14 
VDC),  cable  connections  are  normal  and  the  system 
is ready to operate. 

 

If  the  needle  moves  to  the  red  area  of  the  fault 
monitor meter, output is zero and a short is indicated. 
Short could be located in amplifier, cable connectors, 
or power unit. 

 

If the pointer moves into the yellow area of the fault 
monitor  meter,  an  open  circuit  is  indicated  with  full 
power  supply  voltage.  An  open  circuit  could  be  the 
result  of  a  faulty  amplifier,  an  open  cable  or  open 
connectors. 

 

Allow the sensor to stabilize for about one minute. A 
signal drift may occur when cable is connected to the 
readout instrument. 

 

This  drift  occurs  during  charging  of  the  coupling 
capacitor in the power unit.  The signal  will stabilize 
in several minutes. 
 

4.0 

CALIBRATION 

 

Series  138  are  calibrated  dynamically  using  a  drop 
weight  tester.  For  best  accuracy,  use  the  calibration 
certificate supplied. Factory recalibration is available 
for a nominal charge. 

 

5.0 

POLARITY 

 

This  sensor  series  produces  a  positive-going  output 
voltage  for  increasing  pressure  input.  See  the 
specification sheet for polarity of the sensor. 

 

6.0 

LOW-FREQUENCY RESPONSE 

 

The  low-frequency  response  of  an  ICP

®

  system  is 

determined by: 

 

l.)  The discharge time constant of the sensors. 

 

2.)  If  AC-coupled  at  the  power      unit,      the 
coupling time constant. 

 

Содержание ICP 138A50

Страница 1: ...ensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail in...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...d over a single conductor coaxial cable with a shield serving as signal return Special low noise cables are not required Series 138 are used by the military for underwater explosive testing commercial...

Страница 7: ...high shock can flex the cable connector joint enough to move the center pin of the 10 32 cable connector This can cause noise on the output signal Protecting the connection from shock may be required...

Страница 8: ...nt ranges and for driving long cables to several thousand feet use the higher current up to 20 mA maximum Line impedance matching may be necessary to produce flat frequency response over long hundreds...

Страница 9: ...or l Keep connections clean and dry 2 Do not exceed specified maximum range 3 Do not subject sensor to temperatures exceeding 100 F 4 Do not apply voltage to sensor without current limiting diodes or...

Страница 10: ...8 to14 VDC Physical SensingElement Tourmaline Tourmaline HousingMaterial Stainless Steel Stainless Steel Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 75 oz 21 gm 2 All specific...

Страница 11: ......

Отзывы: