The lowest amplitude a healthy human ear can detect is 20 millionths of a Pascal (20

Pa). Another scale 

more commonly used to describe sound pressure level is the decibel (dB). To convert the output voltage 
signal (measured in Vrms) to sound pressure level (measured in decibels), use the following formula: 

dB

P

S

V

SPL

ref

rms





Log

20

 

where 

is the sensitivity of the microphone in mV/Pa, and 

P

ref

 is the reference pressure in air, which is 

20  x  10

-6

  Pa.  The  decibel  scale  is  logarithmic  and  more  closely  matches  the  response  reactions  of  the 

human ear to the pressure fluctuations. Some examples of typical sound pressure levels are shown in the 
table below. 

dB 

Pressure 

Example 

0.00002 Pa 

Threshold of Hearing 

60 

0.02 Pa 

Business Office 

80 

0.2 Pa 

Shop Noise 

94 

1 Pa 

Large Truck 

100 

2 Pa 

Jackhammer 

120 

20 Pa 

Airplane Take-Off 

140 

200 Pa 

Threshold of Pain 

At  very  high-pressure  amplitudes,  the  microphone’s  diaphragm  will  start  to  distort  the  measured  sound 
pressure.  The  maximum  rated  sound  pressure  level  of  a  microphone  is  expressed  as  the  amplitude  at 
which  the  Total  Harmonic  Distortion  (THD)  reaches  a  specified  amount,  typically  3%  THD.  It  is  im-
portant to note that the highest sound pressure level that can be measured with a microphone system may 
be limited by other parts of the system,  such as  the preamplifier, signal  conditioner, or instrumentation. 
Conversely, the Cartridge Thermal Noise (CTN) specification provides the lowest measurable sound pres-
sure level that can be detected above the electrical noise inherent within the microphone. 

Temperature and pressure will  affect  the sensitivity of  a microphone  but  these  effects  can be accounted 
for  by  adjusting  the  recorded  microphone  values  using  the  coefficients  specified  for  each  microphone 
model. Simply calculate  the difference in  temperature and  ambient  pressure from  those  recorded  during 
calibration  and  multiply  this  value  by  the  proper  coefficient  to  determine  the  sensitivity  offset.  In  most 
cases, the offset will be very small and is therefore unnecessary. 

5.0

 

Accessories 

Transducer Electronics Data Sheet (TEDS) 

Preamplifiers can incorporate TEDS devices, which have a built in read/write memory that contains rele-
vant  information  about  the  sensor  and  its  use.  Information  includes  manufacturer  name,  model  number, 
serial number, sensitivity, etc. The operation of TEDS devices is defined by IEEE P1451.4. A TEDS mi-
crophone system includes a microphone mated with a preamplifier that contains the TEDS memory pro-
grammed with both the microphone and preamplifier information. This is particularly helpful when using 
large channel count systems and array set-up. 

A-Weight Inline Filter 

The inline filter is  used in  conjunction  with  an ICP

®

  preamplifier to  provide A-weighting for acoustical 

measurements. The filter’s frequency response is weighted according to the A-weighting filter portion of 
both ANSI S1.4 and IEC 60651 Type 1 Sound Level Meter Standards. ICP

®

 signal conditioners of 4 mA 

or greater are recommended when using inline filters. 

Windscreens and Nose cones 

Wind  induced noise can be reduced by using a  windscreen.  This  can offer some protection against dust 
particles  and  mechanical  damage.  In  the  presence  of  high-speed  airflow  from  a  well-defined  direction, 
such  as  wind  tunnels,  a  nose  cone  is  recommended.  When  using  windscreens  and  noise  cones,  the  fre-
quency response of the microphone will be attenuated at higher frequencies. 

Содержание 426A14

Страница 1: ...Preamplifier Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 098...

Страница 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Страница 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...he instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument Standards th...

Страница 7: ...hone System Prepolarized Microphone System Phantom Powered Microphone System The phantom powered preamp may be used with either a or a microphone cartridge by using the adapter provided All IEC 61094...

Страница 8: ...icrophone but these effects can be accounted for by adjusting the recorded microphone values using the coefficients specified for each microphone model Simply calculate the difference in temperature a...

Страница 9: ...eamplifier should provide stable and accurate results for years to come Dust rain oil moisture or exposure to extreme temperatures may adversely affect the microphone and preamplifier s performance If...

Страница 10: ...o charge You may also choose to have your purchase price refunded See the supplemental sheet con tained with this manual for information on our service repair and return policies procedures and instru...

Страница 11: ...eel Stainless Steel Size Diameter x Length 0 79 in x 6 50 in 20 1 mm x 165 1 mm Weight 6 23 oz 176 5 gm 1 Electrical Connector 3 Pin XLR 3 Pin XLR Mounting Thread Microphone to Preamplifier 0 2244 60...

Страница 12: ...IFIER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 MCK 3 27 17 JDM 3 27 17 MJN 3 27 17 66288 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any repr...

Отзывы: