background image

Model Number

357B61

CHARGE OUTPUT ACCELEROMETER

Revision: K

ECN #: 46956

Performance

ENGLISH

SI

Sensitivity(± 10 %)

10 pC/g

1.02 pC/(m/s²)

Measurement Range

± 1000 g pk

± 9810 m/s² pk

Frequency Range(± 5 %)

5 kHz

5 kHz

[2]

Resonant Frequency

 24 kHz

 24 kHz

Non-Linearity

 1 %

 1 %

[3]

Transverse Sensitivity

 3 %

 3 %

Environmental

Overload Limit(Shock)

± 5000 g pk

± 49,050 m/s² pk

Temperature Range

-65 to +900 °F

-54 to +482 °C

Temperature Range

-65 to +900 °F

-54 to +482 °C

Temperature Response

See Graph

See Graph

[1]

Base Strain Sensitivity

0.007 g/µ

ε

0.07 (m/s²)/µ

ε

[1]

Radiation Exposure Limit(Integrated Neutron Flux)

1 E10 N/cm²

1 E10 N/cm²

Radiation Exposure Limit

(Integrated Gamma Flux)

1 E8 rad

1 E8 rad

Electrical

Capacitance

630 pF

630 pF

[1]

Insulation Resistance(900° F [482° C])

>100 kohm

>100 kohm

Insulation Resistance

(at 70° F [21°C])

>10

8

 Ohm

>10

8

 Ohm

Physical

Sensing Element

Ceramic

Ceramic

Sensing Geometry

Compression

Compression

Housing Material

Inconel

Inconel

Sealing

Hermetic

Hermetic

Size (Hex x Height)

5/8 in x 1.0 in

5/8 in x 25.4 mm

Weight

1.1 oz

30 gm

[1]

Electrical Connector

10-32 Coaxial Jack

10-32 Coaxial Jack

Electrical Connection Position

Side

Side

[4]

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

is a registered trademark of PCB Group, Inc. 

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

P

- Positive Output Polarity

Output Polarity

Positive

Positive

NOTES:

[1] Typical.
[2] Low frequency response is determined by external signal conditioning electronics.
[3] Zero-based, least-squares, straight line method.
[4] See PCB Declaration of Conformance PS081 for details.

SUPPLIED ACCESSORIES: 

Model 023A10 Hardline cable, 10-ft, 10-32 plug to 10-32 jack (1)
Model 081A107 Mounting stud (10-32 to 10-32) (1)
Model ACS-1 NIST traceable frequency response (10 Hz to upper 5% point). (1)
Model M081A107 Metric mounting stud (10-32 to M6 x 0.75) (1)

Entered: LK

Engineer: BAM

Sales: WDC

Approved: BAM

Spec Number:

Date: 8/8/2017

Date: 8/8/2017

Date: 8/8/2017

Date: 8/8/2017

18161 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

Содержание 357B61/NC

Страница 1: ...celerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E m...

Страница 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Страница 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...F 482 C and extreme application up to 100 F 650 C for UHT 12 These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high tem...

Страница 7: ...urface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the moun...

Страница 8: ...or most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are que...

Страница 9: ...o a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitiv...

Страница 10: ...dure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes asso...

Страница 11: ...Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the tes...

Страница 12: ...s Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz...

Страница 13: ...ector 10 32 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Side Side 4 All specifications are at room temperature unless otherwise specified In the interest of constant product improve...

Страница 14: ...ONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION 218 5 54 X 300 7 62 MIN 1 4 28 UNF 2B X 200 5 08 MIN F M6 X 0 75 M6 X 1 00 M...

Страница 15: ......