PCB Piezotronics 2308-02A Скачать руководство пользователя страница 8

 

 

 

 

 

REACTION TORQUE SENSOR OPERATION MANUAL 

 

 

 

Figure 4 - Axis and Sense Nomenclature for PCB Reaction 

Torque Sensors

 

 
The  principal  axis  of  a  transducer  is  normally  the 
M

Z

 axis. The M

Z

 axis will also be the axis of radial 

symmetry or axis of rotation. In the event there is no 
clearly  defined  axis,  the  following  preference 
system will be used z,x,y.  
 

2.0  SAFETY INFORMATION 

 
Failure  of  the  sensor  structure  or  fasteners  used  in 
its  installation  may  cause  personal  injury  and 
equipment  damage.    To  prevent  structure  failure, 
review  the  extraneous  load  limits  listed  on  the 
specification  sheet  supplied  with  this  manual.  
Extraneous  load  limits  are  extraneous  side  force, 

thrust  and  bending  moment  that  may  be  applied 
without  electrical  or  mechanical  damage  to  the 
torque  sensor.    Do  not  exceed  moment  (W  x  S)  or 
shear  (W)  whichever  attained  first.  Measurement 
inaccuracy and structure damage may result.  Install 
the  sensor  in  a  manner  that  minimizes  these  loads.  
In  addition,  review  data  from  manufacturers  of 
fixtures  and  fasteners  used  in  the  sensors 
installation  to  determine  if  failure  might  occur  due 
to these loads. 
 

3.0  MECHANICAL INSTALLATION 

 
Refer  to  the  outline  drawing  supplied  with  this 
manual  for  specific  outline  dimensions  and 
installation  details  for  your  particular  model.  The 
specification  is  also  included  to  provide  details  of 
the sensor

s characteristic properties.   

 

Shown  below  are  some  of  the  standard  reaction 
torque  sensor  configurations  offered  by  the 
Force/Torque Division. 
 
 
 

 

Figure 5 

 Reaction Torque Sensor Configurations 

 

Содержание 2308-02A

Страница 1: ...UE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ity regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme shock load or other environmental influence or prior to any critical test PCB Piezotronics maintains an ISO 9001 certified metrology laboratory and offers calibration services which are accredited by A...

Страница 3: ...urchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers should direct all inquiries to their local distributor or sales office A complete list of distributors and offices can be found at www pcb com Customers within the United States may contact their local s...

Страница 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Страница 6: ...0 Operation 5 8 0 Troubleshooting 6 9 0 Maintenance 6 LIST OF ILLUSTRATIONS Figure Page 1 Flange Mount Reaction Torque Sensor 2 2 Wheatstone Bridge Circuit 2 3 Western Regional Strain Gage Committee Wiring Code 2 4 Axis and Sense Nomenclature 3 5 Reaction Torque Sensor Configurations 3 6 Shunt Resistor Locations 4 7 Reaction Torque Sensor System Configurations 5 MANUAL NUMBER 18622 MANUAL REVISION...

Страница 7: ...ring Code Refer to the wiring drawing included with this manual for specific wiring of the supplied torque sensor The gages are bonded to the sensor s structure Typically a regulated DC or AC excitation is applied between A and D of the bridge When torque is applied to the sensor the Wheatstone Bridge becomes unbalanced causing an output voltage between B and C which is proportional to the applied...

Страница 8: ...ust and bending moment that may be applied without electrical or mechanical damage to the torque sensor Do not exceed moment W x S or shear W whichever attained first Measurement inaccuracy and structure damage may result Install the sensor in a manner that minimizes these loads In addition review data from manufacturers of fixtures and fasteners used in the sensors installation to determine if fa...

Страница 9: ...junction with a strain gage based transducer without exposing the transducer to known traceable physical input values If required adjustments can then be made to the signal conditioner to insure accurate measurement results The strain gage bridge is in balance when the host mechanical structure is unloaded and unstressed As the host structure diaphragm bending beam shear beam column etc is loaded ...

Страница 10: ...alibration of a signal conditioner and transducer between calibrations of known applied traceable mechanical input values Consequently most all strain gage transducer manufacturers collect and supply shunt calibration data along with a shunt calibration resistor as a standard feature 7 0 OPERATION Figure 7 Reaction Torque Sensor System Configurations Typical Reaction Torque Sensor System Configura...

Страница 11: ...s not typically measure or compensate for errors due to extraneous loads If the presences of these extraneous loads are known the user should request the transducer manufacturer to run a special test at extra cost to define and quantify the extraneous load errors These errors are defined as cross talk errors Error Analysis The Force Torque Division typically supplies accuracy information on its pr...

Страница 12: ...ANUAL 7 Caution should be observed to insure that liquids are not permitted to migrate into devices that are not hermetically sealed Such devices should only be wiped with a damp cloth and never submerged or have liquids poured on them ...

Страница 13: ......

Страница 14: ...18 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 25759 PCB Load Torque A Division of PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Load and Torque REVISIONS REV DESCRIPTION ECO A REDRAWN UPDATED 02 23 18 PTE 47634 n 13 3 2 3X THRU EQUALLY SPACED ON A n1 687 n4...

Отзывы: