background image

 

 

 

 

 

ICP

 FORCE SENSOR OPERATION MANUAL 

 

 7 

traceable reference sensor.  Generally, a preload of 20% (full 

scale operating range of the force ring) is applied before  

recording of measurement data.  Allow the static component of 

the signal to discharge before calibration. 

 

 

3-component force sensors must be preloaded to achieve proper 

operation, particularly for the shear x-, and y-axis. This preload 

provides the sensing crystals with the compressive loading 

required to achieve an output in response to shear direction 

input forces.  The recommended applied preload in the z-axis 

for 3-component force sensors is 10 times the  desired 

measurement range in the x or y-axis.  This higher level z-axis 

preload is required as the resultant output in the x and y axis is 

a result of friction generated on the internal crystals specifically 

cut to provide output in the x or y axis.   

 

  As an example, to maximize the sensors output in all three 

(3ea) axis of a 1000 lb (4500 N) triax load cell, a preload of 

5000 lbs (22K) should be applied,  This will enable 

measurements in the x and y axis to be 500 lbs (2225 N).    

 

Please refer to the specific model specifications for the 

recommended preload. A preload chart in Figure 14  is also 

provided for quick reference.   

 

As with force rings, the sensitivity achieved from a 3-

component force sensor is dependent upon the applied preload 

and the elasticity characteristics of the mounting bolt or stud 

used.  If the unit is to be installed with a stud or bolt other than 

the supplied elastic, beryllium copper stud, a calibration using 

the actual mounting hardware is recommended to be performed.  

Errors in sensitivity of up to 50% can result by utilizing studs 

or bolts of different materials. 

 

When installing ICP

  ring and 3-component type sensors, a 

PCB signal conditioner with at least one channel of DC 

coupling capability is recommended to properly monitor sensor 

output voltage as it corresponds to the desired preload.  A DC-

coupled signal conditioner will provide a longer system 

discharge time constant, which will result in slower signal 

decay.  When used with a DVM or similar readout device, the 

installer can monitor the sensor output voltage directly for 

correct preloading. 

 

Monitor the output from the Z-axis connector when preloading 

3-component type sensors.  These sensors require a preload in 

the Z-axis that is 10 times their shear range.  Some models 

require this to be accomplished in steps, not to exceed the usable 

voltage.  To prevent 

clipping

” 

of the signal, increments should 

not exceed 10 VDC. 

 

  

 

 

Figure 14 - ICP

 Force Sensor Preload Requirements 

 

WORKING RANGE, PRELOAD, AND MAXIMUM 

LOAD RELATIONSHIP IN RING STYLE and Triaxial 

FORCE SENSORS 

 

The 

Working Range

 is the ideal dynamic working load that 

may be applied to a sensor during operation.  In most sensors, 

the product of the working range and the sensor sensitivity 

will provide a 5 Volt output, following the equation;   

5Volts = range x sensitivity.   

 

The 

Sensor Preload

 is the load applied to the sensor before 

the sensor is used in an operation.  In ring and triax models, 

preload is essential to match PCB’s calibrated sensitivity as 

well it assure sensor linearity at the lower measurement range. 

 

The 

Maximum Load

 is the dynamic load that may be applied 

before the sensor approaches physical damage.  In some 

sensors this value is a result of a mechanical limitation.  In 

ICP® models this may be an electrical limitation (applying an 

excessive load under sudden dynamic condition outside the 

specified range may damage the internal electronic circuitry). 

 

With most sensors, the specified dynamic working range and 

maximum compression is riding ON TOP OF the applied 

preload.   As an example, triax force sensor Model 260A01 

has a specified preload of 5000 lbs (22kN), a working range of 

1000 lb (4500N), and a maximum compression range of 1320 

lbs (6000 N). To provide the best linear response of the sensor, 

a 5000 lbs (22kN) preload should be loaded on it.  From there 

one may take dynamic measurements through the entire 1000 

lb (4500N) working range of the sensor.   Dynamically one 

should not take measurements above 1320 lbs (6000) as this 

total load value approaches physical damage to  the sensor. 

 

 

 

Force Ring 

Models 

Preload 

(lbf)

 

Incremental 

Steps 

Sensitivity 

(mV/lbf)

 

Step 

Increment  

(mV) 

 

201B01 

60 

500 

10,000 

201B02 

100 

50 

5,000 

201B03 

200 

10 

2,000 

201B04 

400 

2,000 

201B05 

1,000 

1,000 

202B 

2,000 

0.5 

1,000 

203B 

4,000 

0.25 

1,000 

204C 

8,000 

0.12 

960 

205C 

12,000 

0.08 

960 

206C 

16,000 

0.06 

960 

207C 

33,750 

0.05 

1,688 

3-Component Models 

260A01 

5,000 

2.5 

6,250 

260A02 

10,000 

2.5 

8,333 

260A03 

40,000 

0.25 

10,000 

Содержание 209C01

Страница 1: ...DTC 2 2 lb comp 2200 mV lb Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 6: ...stallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description of the various sensor series available recommended mounting procedures operation and recommended calibration In addition to PCB s line of ICP sensors each ICP sensor series outlined has corresponding ...

Страница 7: ...e ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Installation Methods GENERAL PURPOSE RADIAL Model 208C01 C05 General Purpose Sensors are designed to measure compression and impact forces from a fraction of a lb N to 5 000 lbs to 22 24 kN Tension force...

Страница 8: ... simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the specific plane is obtained when a force is applied axially to the specific x y or z plane Sensor output from loads applied in the alternate transverse planes considered channel cross talk is negligible ...

Страница 9: ...returning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for compression and impact force measurements in materials testing applications Smooth cylindrical housings and curved impact caps avoid cutting through specimens This enables measurements to be taken to de...

Страница 10: ...ot facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surfaces are flat and parallel Non parallel surfaces could place unwanted stress on internal components leading to premature sensor failure The protective cap should remain on the electrical connector durin...

Страница 11: ...the left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total force for the beryllium copper stud supplied with the sensor and up to 50 for steel studs This typical installation setup is used by PCB during standard calibrations of force ring models 201B01 through ...

Страница 12: ...nitor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accomplished in steps not to exceed the usable voltage To prevent clipping of the signal increments should not exceed 10 VDC Figure 14 ICP Force Sensor Preload Requirements WORKING RANGE PRELOAD AND MAXIM...

Страница 13: ... hundreds of feet meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent circuit contained in all PCB signal conditioners See Guide G 0001B for powering and signal conditioning information pertaining to all ICP instrumentation Most of the line powered signal conditioners ...

Страница 14: ...tant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents the decay rate of an input signal One DTC represents the amount of time taken for the signal to decay to 37 of the initial peak value As illustrated in Figure 16 this is an exponential decay Approxim...

Страница 15: ...turned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specification sheet B Do not subject the sensor to temperatures exceeding that of the specification normally 250 F 121 C C Do not apply voltage to the sensor without current limiting diodes or other current pr...

Страница 16: ...Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Electrical Connection Position Side Side Mounting Thread 10 32 Female No Metric Equivalent All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc M Metric Mount Supplied ...

Страница 17: ......

Отзывы: