PCB Piezotronics 1404-13A Скачать руководство пользователя страница 13

 

PCB Load & Torque, Inc.   

 Toll-Free in USA 866-684-7107   

 716-684-0001   

 www.pcbloadtorque.com

 

 

 

GENERAL PURPOSE LOW PROFILE LOAD CELL OPERATION MANUAL 

 

8

 

are  not  allowed  to  migrate  into  devices  that  are  not 
hermetically sealed.  Such devices should only be wiped with 
a damp cloth, and never be submerged or have liquids poured 
on them.  Never use a pressure washer on the load cells. 
 
General purpose low profile load cells,  when  mounted on the 
factory  supplied  base,  are  barometric  compensated  through  a 
small  port  located  in  one  of  the  spanner  wrench  holes  on  the 
base. 
 
The  bolts  used  to  install  bases  at  the  factory  have  been 
tightened  to  specifications  to  prevent  loosening  under  normal 
usage.  However, if the base has been removed for any reason, 
or the load cell has been mounted on a custom fixture without 
the base, it is a good idea to periodically test the mounting bolt 
tightness using a calibrated torque or click wrench. 
 
 

11.0 TROUBLE SHOOTING 

 
Proper performance of a load cell requires careful attention to 
both electrical and mechanical aspects of the measurement 
system.  A basic understanding of the electrical and 
mechanical installation requirements is recommended. 

11.1 Mechanical Trouble Shooting 

A mechanical checklist includes: 
 

1.

 

Check for proper installation of load cell. 
 

2.

 

Check for properly tightened bolts. 

11.2 Electrical Trouble Shooting 

An electrical checklist should start with: 
 

1.

 

Check cables for proper wiring. 

 
2.

 

Inspect for loose or dirty electrical connections. 

 
3.

 

Check for improper shield grounds. 

 
4.

 

Check  for  proper  grounding  of  the  structure  that  the 
load cell is mounted on. 

 
5.

 

Check  the  signal  conditioning  electronics  for  proper 
setup. 

 
6.

 

Check 

the 

insulation 

resistance 

of 

shielded 

conductors for short circuits. 

 
7.

 

Check  isolation  resistance,  load  cell  flexure  to 
conductors. 

 
8.

 

Check  load  cell  bridge  resistances,  (A-D)  excitation 
and (B-C) the signal leads. 

 

9.

 

Check bridge balance. 

 
10.

 

Keep a record of your observations, correct problems, 
or contact PCB factory for assistance. 

11.2.1 Estimating Bridge Balance using an Ohm Meter 

 
A  load  cell  that  has  been  severely  overloaded  will  exhibit  a 
significant  zero  offset  in  the  Wheatstone  bridge.    It  can  be 
useful  to  estimate  the  bridge  unbalance  using  a  digital  ohm 
meter with resolution of at least 0.1 ohm. 
 
Following  the  Western  Region  wiring  diagram  measure  the 
resistances for each leg of the bridge, i.e.:  R

AB

, R

AC

, R

DB

, and 

R

DC

 
 
 
 
 
 
 
 
 
 
 
 

1.

 

Estimate the Zero Offset (mV/V): 

 

Zero Offset = 1.4 x (R

AC

 - R

AB

 + R

DB

 - R

CD

)  

 

2.

 

Estimate the Bridge Unbalance (% Full Scale): 

 

Bridge Unbalance = 100 x (Zero Offset / Output

FS

Where: 

Output

FS

 = Full Scale output of the load cell (mV/V) 

 
A Zero Offset of greater than 10% indicates probable overload 
or  possible  fatigue  damage.    A  load  cell  that  has  been 
significantly overloaded enough to create a large zero offset is 
not  repairable.  In  some  instances  it  may  be  possible  to 
temporarily  continue  to  use  the  load  cell  by  attempting  to 
balance the bridge through use of external resistors. However, 
the user should expect that such a load cell no longer will meet 
the 

performance 

parameters 

related 

to 

its 

original 

specifications. (User beware!!) 

 

12.0 CALIBRATION / REPAIR SERVICES 

 
PCB  Load  &  Torque  offers  calibration  and  repair  services. 
The  PCB  Calibration  Laboratory  in  Farmington  Hills, 
Michigan is A2LA Accredited per ISO/IEC 17025. 
 
 

 

R

DB

 

R

CD

 

R

AB

 

R

AC

 

Figure 17

 

– Wheatstone Bridge 

Содержание 1404-13A

Страница 1: ...OFILE LOAD CELL Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ... repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment calibration cycles are typically established by the users own quality regimen When in doubt about a calibration cycle a good rule...

Страница 3: ...ory customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and mounting hardware are not covered by warranty Mechanical damage...

Страница 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Страница 6: ...r 4 5 0 ELECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6 6 1 2 Hysteresis 6 6 1 3 Best Fit Output 6 6 1 4 Strain Gage Measurements 6 6 1 5 Shunt Calibration Standard Resistor 6 6 1 6 Static Error Band SEB 6 7 0 MOMENT COMPENSATION 6 8 0 FATIGUE OVERLOAD 6 ...

Страница 7: ...evaluated for potential failure modes of the fixtures It is important to use threaded fasteners properly when assembling test fixtures used with the load cells Fixtures with inadequately designed bolted joints or improperly tightened threaded fasteners can fail before the parts under test have completed the planned fatigue test cycles 3 0 OVERVIEW General purpose low profile load cells come in a v...

Страница 8: ...r modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is to be mounted to a custom base or fixture it is necessary to consider the following 1 The thickness of the new base or fixture must not be any thinner than the original base supplied with the load cell Materials used must have the same temperature coefficient of expansion for example 4140 Steel 2 The mountin...

Страница 9: ...of Connector Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not break or loosen any wire connections 3 Slide the connector protector 54823 02 behind the connector as shown in Figure 8 4 Apply a drop Loctite Threadlocker Blue 242 to the threads of the 4 40 x 3 8 s...

Страница 10: ... A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same potential it may be necessary to provide a secure structure ground perhaps by driving a copper rod and attaching a ground strap 6 0 CALIBRATION Every ge...

Страница 11: ...s into the load cell structure The strain gage bridges on general purpose low profile load cells are adjusted such that the error due to eccentric loading is less than 0 25 per inch of offset for 1200 series load cells and 0 1 per inch of offset for fatigue rated 1400 series load cells Side loads moments and torque must be limited to avoid electrical or mechanical damage to the load cell The allow...

Страница 12: ...rces on the load cell If possible remove all loads 2 Power up the host signal conditioner and connect it to the load cell via appropriate cable and allow for a 30 minute warm up 3 Set the load indicator display to read exactly 00 000 4 Connect the shunt resistor to the terminals specified in the calibration certificate and adjust the span or gain until the display reads the force value stated on t...

Страница 13: ...he load cell is mounted on 5 Check the signal conditioning electronics for proper setup 6 Check the insulation resistance of shielded conductors for short circuits 7 Check isolation resistance load cell flexure to conductors 8 Check load cell bridge resistances A D excitation and B C the signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB facto...

Страница 14: ...rder Please request a return material authorization RMA before sending a load cell back to the factory for any reason For calibration services if possible a copy of the purchase order covering the requested services should be included with the returned load cell 13 0 WARRANTY Standard warranty on general purpose low profile load cells covers parts and workmanship For full details refer to the Warr...

Страница 15: ...nted Steel Sensing Element Strain Gage Strain Gage Mounting Torque 1400 lbf in 158 Nm Deflection at Full Scale Capacity 0 001 in 0 025 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice 1404 13B PC04E 10 6P Connector 1404 13...

Страница 16: ...RIDGE 1 2 SUPPLIED WITH MATING FASTENERS WIRING CODE FOR T SERIES LOAD CELLS PIN A EXC PIN B SIG PIN C SIG PIN D EXC PIN E TEDS DATA PIN F TEDS GROUND 1204 M1204 1404 AND M1404 SERIES BASES n2 59 65 8 n5 125 130 2 B C 3 8 24 UNF 2B x 88 12X 1 45 n6 06 153 9 1 75 44 5 94 23 9 37 9 4 n2 25 57 2 030 0 8 THREAD x1 37 34 8 STANDARD LOAD CELL 1 1 4 12UNF 3B METRIC M LOAD CELL M33X2 0 4H0 4X Ø 313 7 94 x...

Отзывы: