Drawing Number: 21096 
Revision: NR

 

 

SERIES 134A20 TOURMALINE PRESSURE BAR GENERAL OPERATION MANUAL

 

Connect  the  Series  134A20  Tourmaline  Pressure 
Bar  with  In-Line  Amplifier  to  a  PCB  power 
conditioner.  Connect  the  output  of  the  power 
conditioner to an appropriate readout instrument. 

 

 

Tie  the  three-inch,  low-noise  cable  to  a  rigid 
structure  to  prevent  excessive  motion  and  noise. 
Allow for strain relief. 
 

Caution

 

If  the  tourmaline  pressure  bar  or  the  three-inch 
cable  become  disconnected  from  the  in-line 
amplifier,  take  caution  to  momentarily  electrically 
short  out  the  cable  and  tourmaline  pressure  bar 
before  reconnecting  the  cable  to  the  in-line 
amplifier.  This  prevents  any  unusual  high  charge 
buildup  from  damaging  the  input  stage  of  the  in-
line amplifier. 
 
Black  vinyl  tape  is  supplied  on  the  sensing  end  of 
the tourmaline bar to dampen heat transfer into the 
sensing  crystal.  This  tape  can  be  removed  and 
replaced if damaged. 

4.0 

OPERATION 

 
Switch  on  the  (PCB)  power  conditioner  and 
observe the bias monitoring voltmeter (or LED) on 
the front panel. 
 
If the indicator reads proper bias (green area on the 
meter,  approximately  11  volts),  the  connections 
from the Series 402A In-Line Amplifier are correct 
and the in-line amplifier is operational. If the meter 
reads  in  the  red  or  yellow  areas,  the  system  is 
shorted  or  open.  The  short  or  open  circuit  may  be 
in the in-line amplifier, the cable connecting the in-
line  amplifier  to  the  power  conditioner,  or  in  the 
power conditioner. 

Allow the system to stabilize for a minute.  A signal 
drift  may  occur  when  connecting  the  power 
conditioner  to  the  readout  instrument.  This  should 
stabilize  as  components  in  the  power  conditioner 
and readout instrument charge. 
 
If  long  cables  between  the  Series  134A20  and  the 
power conditioner are used, impedance matching  
of  the  long  cable  and  increase  of  the  constant 
current  in  the  power  conditioner  may  be  required 
because  of  the  high-speed  response  of  the  sensor. 
Check with PCB if there are questions. 
 

5.0 

CALIBRATION 

 
The  tourmaline  pressure  bar  must  be  calibrated

 

dynamically

.  The tourmaline pressure bar does not 

measure static pressures. In addition, application of 
high static pressure to the tourmaline bar may cause 
damage. 
 
Two  methods  are  used  at  PCB  to  calibrate  the 
Series 134A20: a shock tube and a hydraulic pulse 
tester.    The  shock  tube  is  used  primarily  to  check 
the  sub-microsecond  response  of  the  tourmaline 
pressure bar.  
 
The  hydraulic  pulse  tester  is  used  to  generate  a 
known,  short-duration  pressure  pulse  at  five 
different  pressure  levels  to  determine  the  average 
sensitivity of the sensor.  
 
Recalibration  services  are  offered  by  PCB.  In 
addition,  the  shock  tube  and  hydraulic  pulse 
calibrator are offered as standard products by PCB. 

6.0 

USE OF CHARGE MODE 

 

TOURMALINE PRESSURE 

 

BAR

 

 
The  tourmaline  pressure  bar  is  a  charge  output 
piezoelectric sensor that can be used without the in-
line 

source 

follower 

amplifier  for  certain 

applications.  Calibration  information  is  supplied 
showing  (charge)  output  sensitivity  of  the 
tourmaline  bar  alone,  along  with  its  crystal 
capacitance. 

Содержание 134A27

Страница 1: ...fier Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail inf...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ...he 402A17 installation and operating manuals that comprise a Model 134A27 Tourmaline ICP pressure bar 20k psi 0 25 mV psi 0 2 microsecond rise time kit For assistance with the operation of this produc...

Страница 10: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 11: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 12: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 13: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 14: ...es 134A Tourmaline Pressure Bar senses and transduces a pressure shock wave into a high impedance electrostatic charge This charge is sent through the three inch low noise cable into the Series 402A I...

Страница 15: ...r conditioner or in the power conditioner Allow the system to stabilize for a minute A signal drift may occur when connecting the power conditioner to the readout instrument This should stabilize as c...

Страница 16: ...IER The Series 402A In Line Amplifier is used to convert the high impedance output of the tourmaline pressure bar into a low impedance output The in line amplifier may also be used as a unity gain amp...

Страница 17: ...Stainless Steel Diaphragm Epoxy Epoxy Sealing Epoxy Epoxy Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 1 4 oz 39 gm 4 All specifications are at room temperature unless otherwise...

Страница 18: ...fier Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail inf...

Страница 19: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 20: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 21: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 22: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 23: ......

Страница 24: ...he 402A17 installation and operating manuals that comprise a Model 134A27 Tourmaline ICP pressure bar 20k psi 0 25 mV psi 0 2 microsecond rise time kit For assistance with the operation of this produc...

Страница 25: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 26: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 27: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 28: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 29: ......

Страница 30: ...fier Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail inf...

Страница 31: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 32: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 33: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 34: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 35: ......

Отзывы: