PCB Piezotronics 121A21 Скачать руководство пользователя страница 8

 

Drawing Number: 21088 
Revision: NR

 

 

 

 

OPERATION MANUAL FOR 

ICP

 INDUSTRIAL PRESSURE SENSOR 

Series 121 

Models 121A21, 121A22, 121A23, 121A24 
Models 121A31, 121A32, 121A33, 121A34 

 3 

 

 

Typical ICP

 Connection 

 

4.1 

POLARITY

 

 
The  121  Series  Pressure  Sensors  produce  a  positive-
going  output  voltage  in  response  to  pressure  on  the 
diaphragm. 
 

5.0 

CALIBRATION 

 
The  discharge  time  constant  of  the  Series  121  is  too 
short  to  allow  calibration  of  the  sensor  by  static 
means.  The 121 series can, however, be calibrated by 
dynamic  means  using  a  pulse  calibrator  such  as  the 
Model  903B02.    By  this  method,  the  sensor  is 
subjected  to  a  series  of  calibrated  step  changes  in 
which a linear graph can be plotted over the full range 
of the unit. 
 
PCB  provides  recalibration  service  on  request.    If 
such service is needed or desired, consult the factory 
for details. 
 
Four  sensors  in  the  121  Series  (121A21,  121A22, 
121A23, and 121A24)  have discharge time constants 
of  sufficient  duration  (100  secs.)  to  use  dead  weight 
testers for calibration. 
 
Following  is  a  brief  laboratory  procedure  for 
calibrating a pressure sensor using dead weights. 

5.1 

STATIC CALIBRATION PROCEDURE 

FOR TIME CONSTANTS OF MORE THAN 100 
BUT LESS THAN 500 SECONDS 

 
1.    Place  sensor  on  test  fixture  with  coaxial  cable 

connected to oscilloscope. 

 
2.    Apply  known  pressure  to  sensor  diaphragm 

(within range of instrument) using dead weight. 

 
3.  Remove weight and read change in output voltage 

(mV) corresponding to input pressure (psi). 

 
4.    Repeat  Steps  2  through  4  in  20%  increments 

within range of sensor. 

 
5.    Plot  readings  on  x-y graph (voltage on the y-axis 

verses pressure on the x-axis). 

 
6.    The  slope of the line and the sensor sensitivity is 

mV/psi. 

 
 
 

6.0 

PRECAUTIONS 

 
1.   To avoid destroying the built-in amplifier, do not 

apply  a  voltage  to  the  sensor  without  a  current 
limiting device (20 mA maximum) in the line such 
as those incorporated in all PCB units. 

 
2.    Do  not  subject  sensor  to  temperatures  exceeding 

250°F (121°C). 

 
3.  Do not exceed maximum pressure rating. 
 
 
 
 
 
®ICP is a registered trademark of PCB Group 

Содержание 121A21

Страница 1: ...ensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail in...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...121 are designed for permanent installation by Original Equipment Manufacturers in pumps compressors and pipelines for monitoring pressure fluctuations and or surges Series 121A20 Industrial Pressure...

Страница 7: ...able to the sensor and the other to the sensor input jack on an ICP sensor signal conditioner or equivalent Series 121A30 is provided with an integral twisted pair cable red is signal blue is GND This...

Страница 8: ...e time constants of sufficient duration 100 secs to use dead weight testers for calibration Following is a brief laboratory procedure for calibrating a pressure sensor using dead weights 5 1 STATIC CA...

Страница 9: ......

Страница 10: ......

Отзывы: