Drawing Number: 21105 
Revision: NR

 

 

 

 

OPERATION MANUAL FOR 

BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR 

FOR AMMUNITION TESTING 

Model 090B 

 

 3 

 
5. 

Without 

dropping 

pressure, 

immediately 

proceed  to  next  pressure  level,  set pressure and 
record reading. 

 

NOTE: 

For sensors with full scale range of 

40,000  psi  or  less,  calibration  increments  of 
5,000  psi  are  used.    For  sensors  with  full  scale 
range  above  40,000  psi,  increments  of  10,000 
psi are used. 

 

Proceed in this manner to full scale. 

 
6. 

It is important to note that the test pressure must 
always  be  approached  from  below,  i.e.  by 
increasing pressure, never decreasing. 

 
 

Use  only  unfired  cases  since  the  hardness  of 
fired cases may have been altered. 

 

Do not cycle case before calibrating.  Take one 
set  of  readings  on  each  case,  then  discard  the 
case. 

 

If  the  test  pressure  at  any  point  is  exceeded  by 
more than 5%, discard the case and the data and 
start over with another case. 

 

8. 

Repeat procedure for each case to be 
calibrated. 

 

6.0 

ORGANIZING THE DATA 

 
Since  the  family  of  values  of  output  charge  at  each 
calibration point follows a random distribution about 
a mean value, the standard deviation at each point is a 
valid  indication  as  to  the  preciseness  of  the 
calibration at that point. 
 
A  convenient  format  for  tabulating  this  data  and 
calculating the standard deviation is shown in figures 
1 and 2. 
 

7.0  

UTILIZING THE DATA 

 
See  "An  Improved  Technique  for  Utilization  of 
Conformal  Ballistics  Sensor  Calibration  Data", 
attached. 
 

 
 

 
8.0 

OPERATION 

 
Apply  pressure  slowly,  starting  with  the  lowest 
pressure working toward the highest, set the pressure 
to  specific  values  precisely  and  read  corresponding 
voltage  output  at  each  point.    Do  not  dump  pressure 
at each point. 
 
Due  to  the  fact  that  a  certain  amount  of  pressure  is 
necessary to expand the cartridge case, a severe non-
linearity  will  be  noticed  at  the  low  end  of  the 
calibration curve. 
 
A special technique has been developed to handle this 
problem.   Refer to the enclosed paper "An Improved 
Technique  for  Utilization  of  Conformal  Sensor 
Calibration  Data"  for  instructions  in  collecting  and 
using the conformal calibration data. 
 
Also,  read  the  operating  guide  supplied  with  the 
sensor for further pertinent information regarding use 
of the 090B Calibrator. 
 

9.0 

PRESSURE LIMITS 

 
It  is  very  important  that  maximum  pressures  are  not 
exceeded for the safety of operating personnel. 
 
For  pistol  calibrators,  and  22  rimfire,  the  maximum 
pressure is 40,000 psi.  For center fire rifle cartridges, 
the maximum hydraulic pressure is 60,000 psi. 
 
Do not exceed these pressures.  The units were tested 
at  the  factory  only  to  the  pressures  stated  here.  
Exceeding  these  pressures  may  endanger  personnel 
operating this equipment. 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание 117B227

Страница 1: ...PRESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 6: ...mmunition under test to take into account transmissibility characteristics of the cartridge case for each lot of ammunition Because a certain amount of pressure is required to fully obturate the cartridge case in the test chamber a rather severe non linearity is exhibited over the first several thousand psi of input pressure i e until the cartridge case is pressed firmly against the inside chamber...

Страница 7: ...or two percent of full scale is specified for quartz sensors An acceptable instrument is then defined as one whose data points all fall within this error band The sensitivity of an instrument so calibrated is simply the full scale output charge or voltage divided by full scale input pressure where F S output is determined by the straight line as defined above input S F output S F y Sensitivit Eq 1...

Страница 8: ...ance between the cartridge case and the chamber wall Obturation of the cartridge occurs in this region However when obturation is complete the sensor output is then linear with pressure from this point up to maximum rated pressure In the past a common procedure has been to calibrate the conformal at the expected nominal pressure deriving a sensitivity at this point e g point Po Qo in figure 4 The ...

Страница 9: ...he data from 20 consecutive calibration runs performed on a conformal sensor yielded a slope of 04 pC psi and an intercept of 7 500 psi determined by constructing the best fitting straight line as described previously Assume also that average expected peak pressure is 25 000 psi Approximate full scale charge output is pC 1000 pC psi 04 psi 25 000 Set range of charge amplifier output to 1000 pC vol...

Страница 10: ... can be further refined to include a variable voltage source as a bias in series with the output of the charge amplifier This voltage bias can be set to simulate the offset pressure and then the true peak pressure can be read directly See figure 6 on the next page Example Assume a 20 round calibration lot and attendant graph yield a slope of 250 pC psi and an offset of 5 000 psi Proceed as follows...

Страница 11: ...scale pressure to find full scale picocoulombs 40 000 psi x 250 pC psi 10 000 pC Set the charge calibrator for 10 000 pC and apply this signal to the input of the amplifier Now adjust the gain of the oscilloscope or charge amplifier if gain adjust is available to make full scale equal to a selected number of oscilloscope divisions For example 6 Div 40 000 psi Now the peak value can be determined f...

Страница 12: ...Drawing Number 21210 Revision NR AN IMPROVED TECHNIQUE FOR UTILIZATION OF CONFORMAL BALLISTICS SENSOR CALIBRATION DATA 7 Adds offset pressure not available from PCB ...

Страница 13: ...ADAPTOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 14: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 15: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 16: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 17: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 18: ...stment The rotational alignment is automatically controlled by the alignment device incorporated in all 117B Conformal Sensors 4 0 INSTALLATION OF THE CARTRIDGE CASE Refer to Installation Dwg No 090 2 90 It is important that cartridge case selected for calibration be identical in all respects including material to the lot of ammunition undergoing test It is advisable that the test unit be selected...

Страница 19: ...s of torque on this nut to affect a pressure seal 4 Connect 117B to an electrostatic charge amplifier such as PCB Model 462A or 462B52 on long time constant Zero the system electrically by moving front panel toggle to gnd position momentarily 5 0 COLLECTING THE DATA It is important to proceed carefully while performing the calibration to avoid procedures which could lend inherent error to the resu...

Страница 20: ...t for tabulating this data and calculating the standard deviation is shown in figures 1 and 2 7 0 UTILIZING THE DATA See An Improved Technique for Utilization of Conformal Ballistics Sensor Calibration Data attached 8 0 OPERATION Apply pressure slowly starting with the lowest pressure working toward the highest set the pressure to specific values precisely and read corresponding voltage output at ...

Страница 21: ...Drawing Number 21105 Revision NR OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B 4 Figure 1 ...

Страница 22: ...Drawing Number 21105 Revision NR OPERATION MANUAL FOR BALLISTICS PRESSURE SENSOR CALIBRATION ADAPTOR FOR AMMUNITION TESTING Model 090B 5 Figure 2 ...

Страница 23: ...PRESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 24: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 25: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 26: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 27: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 28: ...ngement allows for tolerance in the location of the guide pin hole in an axial direction on the test barrel or calibration adaptor The precise depth adjustment is obtained by the use of the correct thickness spacer selected from a set of 9 spacers of various thicknesses supplied with each sensor Once the proper thickness spacer is found removal and re installation now becomes a routine matter 3 0 ...

Страница 29: ...ect a thinner spacer and remount Once the proper thickness is found for perfect flushness the sensor may be removed and reinstalled using this same spacer and the proper depth will be achieved each time NOTE For best accuracy of results use same charge amplifier for calibration and for actual operation Use long TC for calibration medium or short TC for best drift free operation 4 0 POLARITY Polari...

Страница 30: ...r to allow easy insertion and extraction the case must be expanded slightly by the internal pressure before force can be transmitted to the sensor See the enclosed guide An Improved Technique for Utilization of Conformal Ballistics Sensor Calibration Data for methods of dealing with this topic 6 0 MAINTENANCE It is essential for normal operation of the Model 117B that the insulation resistance be ...

Страница 31: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 4 ...

Страница 32: ...Drawing Number 21086 Revision NR OPERATION MANUAL FOR CONFORMAL BALLISTICS PRESSURE SENSOR Series 117B 5 ...

Страница 33: ...xy Epoxy Electrical Connector 10 32 Coaxial Jack 10 32 Coaxial Jack Weight 0 45 oz 13 gm All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as ...

Страница 34: ...ADAPTOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 35: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 36: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 37: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 38: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 39: ...43 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN B UPDATED ALIGNMENT GUIDE 39992 O RING 2 011 PARKER EMPTY CARTRIDGE CASE 480 RUGER ALIGNMENT GUIDE MODEL 117B227 CONFORMAL BALLISTIC PRESSURE SENSOR ELECTRICAL CONN...

Страница 40: ...PRESSURE SENSOR Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 41: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 42: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 43: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 44: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 45: ...001 23 HOLD THIS DIMENSION TO OBTAIN FLUSH MOUNT 1 2 3 4 5 2 1 4 REVISIONS REV DESCRIPTION DIN B UPDATED 065A19 NOTE 50698 6975 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc 5 236 002 5 99 05 747 18 97 250 6 35 46 11 8 89 22 7 1 63 41 4 3 8 24...

Отзывы: