background image

 

Drawing Number: 21075 
Revision: C 
ECN Number: 47023 

 

OPERATION MANUAL FOR 

QUARTZ PRESSURE SENSORS 

SERIES 102 
SERIES 113 

 

5.0 

POLARITY 

 
The  sensors  in  this  series  produce  a  positive-going 
output voltage for increasing pressure input. 
 

6.0  LOW-FREQUENCY RESPONSE

 

 
The  low-frequency  response  of  an  ICP

®

  system  is 

determined by: 
 
1. 

The discharge T.C. of the sensor  

  

2.        If AC-coupled at power unit, the coupling time 
constant. 
 
Consult  Section  7.0  in  guide  G-0001B  detailed 
explanation  of  low-frequency  characteristics  of  ICP

®

 

instruments. 
 

7.0 

HIGH-FREQUENCY RESPONSE

 

 
Frequency  tailoring  and  the  very  high-natural 
frequency  of  the  sensor  give  an  extremely  wide 
usable 

frequency 

range 

(beyond 

100 

kHz).  

Exceptionally  fast  response  time  (1  µsec)  and  clean, 
virtually  non-resonant  response  to  rapid  step 
functions  are  also  features  of  these  sensors.    As 
mentioned  previously,  the  diaphragm  must  be  flush-
mounted  to  fully  realize  the  high-frequency  response 
capabilities of this series. 

 

8.0 

CALIBRATION

 

 
Piezoelectric  sensors  are  dynamic  devices,  but  static 
calibration  means  can  be  employed if discharge time 
constants  are  sufficiently  long.    Generally,  static 
methods  are  not  employed  below  several  hundred 
seconds time constant.   
 
To  employ  static  means,  direct  couple  the  sensor  to 
the  DVM  readout  using  a  T-Connector  from  the 
sensor  jack  or  use  the  Model  484B  in  the  calibrate 
mode.    Apply  pressure  with  dead  weight  tester  and 
take  readings  quickly.    Release  pressure  after  each 
calibration point. 
 
For  the  shorter  time constant, rapid step functions of 
pressure are generated by a pneumatic pressure pulse 

calibrator  or  dead  weight  tester  and  readout  is  by 
recorder or storage oscilloscope. 
 
PCB  offers  a  complete  calibration  service.    Consult 
factory for details. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

ICP is a registered trademark of PCB Piezotronics 

Содержание 113B26

Страница 1: ... Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ity regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme shock load or other environmental influence or prior to any critical test PCB Piezotronics maintains an ISO 9001 certified metrology laboratory and offers calibration services which are accredited by A...

Страница 3: ...urchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers should direct all inquiries to their local distributor or sales office A complete list of distributors and offices can be found at www pcb com Customers within the United States may contact their local s...

Страница 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Страница 6: ... shows the components of the basic ICP probe i e the piezoelectric element and the ICP source follower amplifier These components are joined together as an inseparable sealed assembly at the factory Disassembly should not be attempted in the field Series 113B2x are in a probe configuration and are installed with a hollow clamp nut with 5 16 24 external threads The housing of these models is at ele...

Страница 7: ...ial in many cases One or more layers may be used across the end of diaphragm and adaptor A silicone rubber coating approximately 010 thick has also been proven effective in many applications General Electric RTV type 106 is recommended Apply the rubber coating to the surface of the diaphragm and allow it to cure in accordance with the manufacturer s instructions If you have ordered the CA option a...

Страница 8: ...s are also features of these sensors As mentioned previously the diaphragm must be flush mounted to fully realize the high frequency response capabilities of this series 8 0 CALIBRATION Piezoelectric sensors are dynamic devices but static calibration means can be employed if discharge time constants are sufficiently long Generally static methods are not employed below several hundred seconds time ...

Страница 9: ......

Страница 10: ......

Отзывы: