Drawing Number: 21113 
Revision: A 

 

OPERATION MANUAL FOR 

ICP

®

 HIGH PRESSURE SENSORS 

Series 108 & 109 

 

Existing Recessed Ports

 

 

If  waveform  distortion  occurs  during  prolonged 
testing,  remove  the  sensor  and  clean  the  residue  as 
illustrated above. 

  

3.2 

PREPARING NEW MOUNTING 
PORTS 

 

Refer  to  the  installation  drawing  provided  in  this 
manual for instructions on mounting hole preparation.  
For  best  results,  do  not  deviate  from  the  steps 
outlined in this drawing. 

 

To assist new mounting port preparation, PCB offers 
the  040B20  Tooling  Kit  (040B21  for  metric mount).  
These  installation  kits  provide  all  of  the  necessary 
tooling  required  to  drill,  ream,  and  tap  the  mounting 
ports for proper installation of the 108, 109, 118 and 
119 Series transducers.   

 

Use  good  machining  practice  in  preparation  of  the 
mounting  port,  paying  particular  attention  to the seal 
surface.    It  is  important  that this surface be perfectly 
smooth  and  free  from  tool  chatter  marks,  nicks  and 
other  imperfections  which  might  cause  leaks  at  high 
pressures. 
 

3.3 

RECESSED MOUNT

 

 

The  recessed  installation  is  shown  in  the  Recessed 
Mount  figure.    This  type  of  mounting  protects  the 
sensor  diaphragm  from  the  effects  of  high-flash 
temperatures  and  particle  impingement  due  to  blast 
effects, thereby prolonging sensor life. 

 

The  recommended  range  of  passage  diameters  is 
0.090 to 0.125 inch (2.29 to 3.18 mm) diameter. 

The  limitation  in  this  type  of  installation  lies  in  the 
frequency-limiting  effects  of  the  passage  due  to  its 
length. 

 

The  passage  behaves  like  an  underdamped  second 
order 

system; 

the  resonant  frequency  being 

determined  by  the  passage  length.    The  length  may 
have a limiting effect on pressure pulse rise time and 
cause  passage  ringing  in  cases  where  the  passage  is 
too long. 
 

Recessed Mount 

 

 

 
The following relationship approximates this resonant 
frequency (f

 r

): 

 

(Hz)

L

4

V

f

r

 

(EQ. 1) 

 
Where:  f

 r

 = Resonant frequency of passage (Hz) 

V = Velocity of sound in air (ft/sec) 

L = Length of column (ft) 

For air at room temperature, (EQ. 1) becomes: 
 

L

00

3

3

f

r

 

 

(EQ. 2) 

 
Where:   L = Passage length (in) 
 
 
 

MODEL 040A 

REAMER 

MODEL 041A 

PILOT BUSHING 

0.010 

MIN 

Содержание 109C12

Страница 1: ... psi accel comp ceramic coated integral diaphragm Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 6: ...ff diaphragm give these sensors a high natural frequency and linearity The microelectronic amplifier converts the high impedance voltage from the quartz package into a low impedance high level output signal See the Technical Information section of our website for more complete coverage of ICP instruments 3 0 INSTALLATION With a 3 8 24 M10x1 0 for metric mount mounting thread and flush diaphragm de...

Страница 7: ... nicks and other imperfections which might cause leaks at high pressures 3 3 RECESSED MOUNT The recessed installation is shown in the Recessed Mount figure This type of mounting protects the sensor diaphragm from the effects of high flash temperatures and particle impingement due to blast effects thereby prolonging sensor life The recommended range of passage diameters is 0 090 to 0 125 inch 2 29 ...

Страница 8: ...se of recessed installation In severe pyrotechnic environments sensor life may be severely limited with flush installation 3 5 FLASH TEMPERATURE EFFECTS The ceramic coating on the diaphragm of these sensors should render the flash thermal effect insignificant in most cases especially when recessed mounted However if more protection from flash thermal effects is required with the recessed mount the...

Страница 9: ...e shows the 108 109 Series connected to a PCB Model 484B06 power unit The 484B06 is a power unit that can be AC or DC coupled In the DC couple mode a level shifting circuit removes the bias level and provides a zero adjust feature at the front panel The system s discharge time constant is determined by the sensor in this mode An AC coupled mode is provided for normal AC coupled operation 4 0 CALIB...

Страница 10: ...w instrument several minutes to warm up and thermally stabilize When output from the power unit is connected to readout equipment a drift in the voltage signal will be noticed this is normal This voltage signal drift is caused by the charging of the coupling capacitor in the power unit and will cease within several minutes 6 0 POLARITY The 108 109 Series are designed to produce a positive output v...

Страница 11: ......

Страница 12: ...ghts in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN F ADDED CERAMIC COATING NOTE 53103 CAUTION DO NOT TOUCH SEAL SURFACE WITH TAP WHEN TAPPING HOLE DIMENSIONS A B TO SUIT USERS REQUIREMENTS KEEP DIMENSION A AS SHORT AS POSSIBLE SEE OPERATING GUIDE FOR OPTIMUM P...

Отзывы: