Accelerometers install and connect according to the procedures described in the specific 
sensor manual.  Accelerometers utilized for impact testing include the 353 Series of 
single axis accelerometers, and the 356 Series of triaxial accelerometers.  For additional 
information regarding sensors to meet your impact testing needs, feel free to contact the 
PCB

®

 Vibration Division customer service group by calling toll-free in the USA 888-

684-0013 (or outside the USA by calling 716-684-0001).   
 
In general, sensors connect to the ICP

®

 power unit at the jack labeled 

XDCR

.  The jack 

labeled 

SCOPE

 is connected to the readout device, FFT analyzer, or computer.  Please 

consult the ICP

®

 power unit operating guide for proper installation and operating 

procedures.  
 
 

4.0 Operation   

Follow the below operation instructions: 
 
1). Assemble the Modal Punch to the desired length using the supplied extension (if 
required) and connector adaptor.  Please reference the above installation procedures. 
 
2). Choose the impact tip which corresponds to your excitation requirements.  In general: 

A.  A stiffer tip is recommended for higher frequency response. 
B.  A softer tip is recommended for improved low frequency response. 

 
3). Connect the Modal Punch to an ICP

®

 power supply or equivalent with a coaxial cable 

(not supplied).  The Modal Punch requires a 10-32 microdot connector plug. 
 
4). Connect the corresponding output cable (not supplied) from your ICP

® 

power supply 

to your readout device.  This step is not necessary if  ICP

® 

 power is built into your 

readout device.  Tighten the cable connectors securely by hand to ensure good electrical 
contact.   
 
5). Install accelerometer(s) in the desired measurement location(s), referring to the 
appropriate operating guides for the accelerometer(s) and power supplies being used. 
 
6). Switch power on and wait a minute for the sensor amplifier to turn on and the 
coupling capacitor to fully charge. 
 
7). Check the ICP

® 

power unit meter for normal operation.  Normal operation is indicated 

when the meter pointer is pointing in the green area.  If the meter reads in the red area, 
look for a shorted cable or connection.  If the meter pointer is pointing in the yellow area, 
look for open cables or connections.  
 
8). When all power unit meters indicate normal operation (meter pointer is pointing in the 
green area), proceed with tests, following all sensor, power unit, and analyzer operating 
instructions.  
 

Содержание 088A

Страница 1: ...Model 084A80 extender Model Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...ng the resultant motion with an accelerometer s Integration of the acceleration signal yields velocity or displacement data for use in computing classical transfer characteristics such as inertance co...

Страница 7: ...41 supplied just below the impact surface An Extension Shaft model 084A80 is also supplied to adjust the Modal Punch to the desired length Additional shafts may also be purchased A foam grip is suppli...

Страница 8: ...for higher frequency response B A softer tip is recommended for improved low frequency response 3 Connect the Modal Punch to an ICP power supply or equivalent with a coaxial cable not supplied The Mo...

Страница 9: ...ling 716 684 0001 5 0 Testing To test the behavior of your structure and to tailor the frequency bandwidth of the impact force follow the below operation instructions 1 Position the Modal Punch on the...

Страница 10: ...perfect this technique Distortion undershoot and oscillation of the analyzer display is caused by ringing of the analyzer s anti aliasing filters This behavior is normal To view the correct impulse wa...

Страница 11: ...of the Modal Punch is designed to provide reliable long term service life Problems may arise when the instrument is exposed to temperatures exceeding the specific range prolonged exposure to moisture...

Страница 12: ...to the user 9 0 CUSTOMER SERVICE PCB Piezotronics guarantees Total Customer Satisfaction If at any time for any reason you are not completely satisfied with any PCB product PCB will repair replace or...

Страница 13: ...ment we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as li...

Страница 14: ......

Отзывы: