5. Input/output signals
The SGI-3, OXI-3 and OXI-3.1 are equipped with an 8-pole DIN connector for connection
to data collection equipment and/or remote control.
Location of pins in the DIN connector seen from the soldering side.
Note: The functions of this connector have been changed in relation to previous
models in some respects
.
5.1. SGI-3 pin connections
pin 1 (+) & pin 2 (-) :
Current output - see also the section “Measuring range” on
page 20.
pin 3 (+) & pin 2 (-) :
“Ready” output:
v
sensor cold (not ready) = 0 VDC
v
sensor warm (ready to measure) = 15 VDC (100mA)
pin 4 (+) & pin 7 (-) :
“Start measuring” input:
v
Stop measuring = 0 VDC
v
Start measuring = 10-32 VDC (10 mA)
pin 6 (+) & pin 5 (-) :
15 VDC supply output, max. 10 mA
pin 5 (-) :
Internally connected to pin 2 (-)
pin 8 (+):
Internally connected to pin 1 (+)
5.2. OXI-3 pin connections
pin 1 (+) & pin 2 (-) :
Current output - see also the section “Measuring range” on
page 20.
pin 3 (+) & pin 2 (-) :
“Ready” output:
v
sensor cold (not ready) = 0 VDC
v
sensor warm (ready to measure) = 15 VDC (100mA)
pin 4 (+) & pin 7 (-) :
“Start measuring” input:
v
Stop measuring = 0 VDC
v
Start measuring = 10-32 VDC (10 mA)
pin 6 (+) & pin 5 (-) :
“Measuring range” output:
v
“%” measuring range = 0 VDC
v
“ppm” measuring range = 15 VDC (100mA)
pin 5 (-) :
Internally connected to pin 2 (-)
pin 8 (+):
Internally connected to pin 1 (+)
5.3. OXI-3.1 pin connections
pin 1 (+) & pin 2 (-) :
Voltage output - see also section “Measuring range” on page
20.
pin 3 (+) & pin 2 (-) :
“Ready” output:
v
sensor cold (not ready) = 0 VDC
PBI-Dansensor A/S
SGI-3/OXI-3/OXI-3.1 User manual
10
PBI-210247-H UK 07/11/2005
6 8
3
5
2
4
1
7