![background image](http://html1.mh-extra.com/html/patterson/sammons-preston-r-flex/sammons-preston-r-flex_service-manual_4016042011.webp)
CORRECTIVE MAINTENANCE
Sammons Preston CPM Unit
Power Supply Replacement Procedure
1. Turn off the power and unplug the CPM unit.
2. Remove the control box cover with a Phillips head screwdriver.
3. Disconnect the circuit board connector and plug connector from the power supply.
4. Using a NON-Magnetic screwdriver, loosen 4 screws and remove power supply from CPM unit.
5. Install new power supply and secure with 4 screws.
6. Reconnect circuit board connector and plug connector to power supply.
7. Reinstall control box cover.
8. Perform auto-calibration procedure on the CPM unit before use.
Disconnect from here
9