U.S Only
Digital Apex Locator
Dual Function Apex Locator
Quick Guide
Verify conductivity between the handpiece and the D.A.L. by connecting
the 0.5
validation cable
to the
D.A.L. Make
contact between
the endo
file
and the lip hook cable. The 0.5 LED on the D.A.L. should light, verifying
conductiov of
the electric
current
between the
D.A.L. and
the endo
file,
allowing accurate measurement of the canal length.
ATTENTION: In or
der to avoid a "false positive" reading if contact with the
soft tissues occurs, the handpiece can be covered with a rubber sleeve.
It is not, or may not serve as an alternative for the rubber dam, which is
mandatory in endo procedur
es. The rubber sleeve is a single use item
and should be disposed after use to prevent cr
oss infection. It cannot be
sterilized. The rubber sleeve is sold seperately.
NOTE: W
orking with
handpiece requir
es a thin
file alter
nately after
the thick
one for
additional cleaning
of the
remnants.
The D.A.L.
will not
be accurate
if the pulp chamber is full of liquid. It should be completly dry.
NOTE: In order to fully pr
epare the canal using the hand piece and r
otary
files, keep
the D.A.L.
"on" and
in measuring
mode (connected),
to ensur
e
that the apex is not perforated.
NOTE: When
starting treatment,
first take
a manual
measurement
using a
manual file.
For further information and
troubleshooting, please consult our full user
manual on the enclosed CD.