5. Cleaning and Drying
7. Sterilization
8. Guarantee
・
Never install/remove the attachment onto/from the handpiece while the
handpiece is in operation.
・
Never operate the handpiece without the attachment.
・
Whenever any abnormalities in performance, sound level, heat level,
function, etc. are sensed, stop the operation and consult the Patterson
Service Center.
・
Never keep the handpiece connected to the tubing connector.
Caution
!
・
Do not use sprays other than PANA SPRAY Plus. Using sprays other
than PANA SPRAY Plus may cause the handpiece to overheat.
・
Hold the spray can upright.
・
Firmly hold the handpiece to prevent it slipping when spray pressure is
applied.
・
Spray lubricant until it expels from the handpiece head.
Caution
!
!
Warning
!
●
Autoclave is recommended for sterilization.
1
. Before sterilization, clean the surface of the handpiece and lubricate the
handpiece.
2
. Lubricate the handpiece and make a trial run.
3
. Remove the attachment from the handpiece.
4
. Put the handpiece into a sterilization pouch. Do not combine it with others.
5
. Place the handpiece in the center or upper shelf of the autoclave chamber.
6
. Autoclave the handpiece at 275
°
F or 135
℃
for 20 minutes.
Patterson guarantees the handpiece for 10 months subject to the observance
of maintenance instructions, cautions, warnings and normal use.
Notes:
●
Patterson reserves the right to analyze and determine the cause of any failure.
●
Guarantee is void if the breakage, accident or any other issues are caused by
the neglect of the daily maintenance, cautions, warnings and common sense
in use.
●
Dismantling or reassembling by any third parties other than the Patterson
Service Center voids guarantee.
Never dismantle/alter the handpiece in any case.
・
Never use dry heat sterilization, cold sterilization, etc.
・
Never autoclave the handpiece together with others in a sterilization
pouch.
・
Never place the handpiece at the bottom of autoclave chamber, as the
temperature at the bottom may rise above the actual setting
temperature.
Caution
●
The handpiece must be cleaned and lubricated after each use and before
autoclaving using the following procedures:
1
. Remove the attachment from the handpiece.
2
. Remove the handpiece from the tubing connector.
Cleaning and Drying the Exterior
Wash the handpiece under running water following the procedures below.
(Water condition: ≤100°F, ≥3.5L/min; water should be of the same quality as
drinking water)
●
Clean the external surfaces of the
handpiece with a soft bristled toothbrush
for 15 seconds or more.
●
Clean around the bur insertion hole for 15
seconds or more, using an interdental
brush (wire thickness 0.7mm) that
conforms to size 4 in ISO 16409.
Under appropriate lighting (natural office
lighting, 500 lx or higher), inspect the
handpiece for blood or other debris. If any
visible debris remains, repeat the process
until the handpiece is visually clean.
After cleaning, if the residual moisture is
present, wipe off with dry cloth or blow it off with compressed air (≤50.75psi)
until there is no moisture in the interior and exterior.
When blowing off with compressed air, cover the handpiece with cloth to
prevent scattering of water. Proceed to 6: Cleaning and Lubrication.
6. Cleaning and Lubrication
Cleaning and Lubricating the Interior
●
Attach a Tip Nozzle to the nozzle of the
PANA SPRAY Plus.
●
Shake the can 3 or 4 times and insert the
Tip Nozzle into the drive air port while
holding the head of the handpiece with a
cloth etc.
●
Hold the handpiece and spray for 2 or 3
seconds until oil comes out of the
handpiece head. Repeat lubrication until foreign material stops coming out of
the tip.
2021-11-00 CACB1100 N
15sec.