5. Cleaning and Drying

7. Sterilization

8. Guarantee

Never install/remove the attachment onto/from the handpiece while the 

handpiece is in operation.

Never operate the handpiece without the attachment.  

Whenever any abnormalities in performance, sound level, heat level, 

function, etc. are sensed, stop the operation and consult the Patterson 

Service Center. 

Never keep the handpiece connected to the tubing connector.

Caution

!

Do not use sprays other than PANA SPRAY Plus. Using sprays other 

than PANA SPRAY Plus may cause the handpiece to overheat.

Hold the spray can upright.

Firmly hold the handpiece to prevent it slipping when spray pressure is 

applied.

Spray lubricant until it expels from the handpiece head.

Caution

!

!

Warning

!

 

Autoclave is recommended for sterilization.    

 

                                              

. Before sterilization, clean the surface of the handpiece and lubricate the 

handpiece. 

. Lubricate the handpiece and make a trial run.

. Remove the attachment from the handpiece.

. Put the handpiece into a sterilization pouch. Do not combine it with others.

. Place the handpiece in the center or upper shelf of the autoclave chamber.

. Autoclave the handpiece at 275

°

F or 135

  for 20 minutes.

Patterson guarantees the handpiece for 10 months subject to the observance 

of maintenance instructions, cautions, warnings and normal use. 

Notes:

 Patterson reserves the right to analyze and determine the cause of any failure.

 Guarantee is void if the breakage, accident or any other issues are caused by 

the neglect of the daily maintenance, cautions, warnings and common sense 

in use.

 Dismantling or reassembling by any third parties other than the Patterson 

Service Center voids guarantee. 

Never dismantle/alter the handpiece in any case.

Never use dry heat sterilization, cold sterilization, etc.

Never autoclave the handpiece together with others in a sterilization 

pouch. 

Never place the handpiece at the bottom of autoclave chamber, as the 

temperature at the bottom may rise above the actual setting 

temperature. 

Caution

 The handpiece must be cleaned and lubricated after each use and before 

autoclaving using the following procedures:  

. Remove the attachment from the handpiece.

. Remove the handpiece from the tubing connector.

Cleaning and Drying the Exterior

Wash the handpiece under running water following the procedures below. 

(Water condition: ≤100°F, ≥3.5L/min; water should be of the same quality as 

drinking water)

 Clean the external surfaces of the 

handpiece with a soft bristled toothbrush 

for 15 seconds or more. 

 Clean around the bur insertion hole for 15 

seconds or more, using an interdental 

brush (wire thickness 0.7mm) that 

conforms to size 4 in ISO 16409.

Under appropriate lighting (natural office 

lighting, 500 lx or higher), inspect the 

handpiece for blood or other debris. If any 

visible debris remains, repeat the process 

until the handpiece is visually clean.

After cleaning, if the residual moisture is 

present, wipe off with dry cloth or blow it off with compressed air (≤50.75psi) 

until there is no moisture in the interior and exterior.

When blowing off with compressed air, cover the handpiece with cloth to 

prevent scattering of water. Proceed to 6: Cleaning and Lubrication.

6. Cleaning and Lubrication

Cleaning and Lubricating the Interior

 Attach a Tip Nozzle to the nozzle of the 

PANA SPRAY Plus.

 Shake the can 3 or 4 times and insert the 

Tip Nozzle into the drive air port while 

holding the head of the handpiece with a 

cloth etc.

 Hold the handpiece and spray for 2 or 3 

seconds until oil comes out of the 

handpiece head. Repeat lubrication until foreign material stops coming out of 

the tip. 

2021-11-00  CACB1100 N

15sec.

Pour une version en français du manuel d'utilisation, veuillez vous rendre sur le 

site Web de Patterson: https://www.pattersondental.com/fr-CA

Manufactured for:

Fabriqué pour :

Patterson Dental Supply, Inc.

1031 Mendota Heights Road

Saint Paul, MN 55120