5.2.1 Water Filling Procedure
Step:
1. Connect the clear tube with the funnel to the port marked FILL on
the upper right side of the service panel
2.Use the supplied deionized water, and pour it slowly into the funnel. Align
the clear tubing with the service panel (see Figure). Stop filling as soon as
the water starts to spill from the Vent port.
3.Plug in the main power supply cable.
4.Connect the desired module to the system.
5.Turn the system on and let it run for five minutes.
6.Turn the system off.
7.Repeat steps 3 and 6 several times, until no more water can be added
(spills out of the elbow right away).
8.Disconnect the Fill/Drain Kit components.
9.Turn the system on; it is ready for operation.