PATHFINDER TSC-20A Скачать руководство пользователя страница 2

packing slip. If any discrepancies exist, be sure to contact 

Pathfinder Instruments 

immediately. Inspect all controllers for damage. 

Check for damaged boxes, scratches on the controller's enclosures and face plates, or any damage that may exist due to improper 
handling. If such a case exists save the shipping carton and shipping material, and contact your shipping agent immediately. 

 
B. Mounting 

To flush mount your controller, first verify that the depth of your cabinet will accommodate the length of your instruments. Also 
verify that where the controllers are to be mounted there are no corrosive gasses present and no vibration, impact, water, or 
extreme temperature exposure. Once verified, follow the diagrams below and cut the correct sized hole from the cabinet's panel; 
the panel should be between 1 and 8 millimeters thick. Insert each unit through the front of the panel. The controller's bezel should 
catch and not feed through the cut out. When the controller is properly fitted in the panel, slide the plastic mounting clamp in place 
and tighten the clamp screws for a firm fit. 

 
C. Wiring Power to Controllers 

AC power is to be connected to the power input terminals located on the back of your 

TSC-20-A 

refer to the wiring diagram 

in the manual or on the side of the controller for the correct terminals. Your power connections should be made with 18 gauge or 
larger insulated wire. A 3 amp fuse should be connected in series with power and your controller to help eliminate any problems 
which could occur due to an over current situation. The 

TSC-20's 

unique power supply circuit, incorporating a free voltage 

transformer accepts line voltages between 100 and 240 VAC; no need to change your wiring to accommodate voltage differences 
within the rated voltage range. 
 

Operation 

A. Connect the probe 

1 .Refer to the wiring diagramand connect the probe to terminals as shown, Red wire to #16, blackwire to # 17, and green wire to # 
18 

B. Keypad Operation 

1. "SEL" key; changes upper display to the first, second and third programming menus, advances the display through the 
menu and sets the programmed information into memory 
2... "/\" up arrow key, increases value displayed in the green (lower) display 
3... "\/" down arrow key, decreases value displayed in the green (lower) display 

C. Programming 

1. 

How to change the main control relay set point 

Press and hold the "/\" up or "\/" down arrow until the correct number appears, the new set point will be active after 5 seconds. 
2. 

How to change the Alarm relay set point 

Hold the “SEL” key for approximately 2 seconds and “AL1” will appear in the upper display, the set point will be in the lower display. 
Press the “SEL” switch once, the value will flash, change the value with the “up” or “down “ keys, and press the “SEL” key again to 
set it in memory. Holding the “SEL” key for approximately 2 seconds will return to normal operation. 
3. 

How to program the main control relay Hysteresis ( Deadband ) 

Hold the “SEL” key for approximately 3 seconds and “HYS” will appear in the upper display, the hystersis amount will appear in the 
lower display, press the “SEL” key once and the amount will flash, change the amount to the desired value with the “up” or “down 
“keys, press “SEL” again to set it in memory. Holding the “SEL” key for approximately 2 seconds will return to normal operation. 
4. 

How to change the main control relay action (Relay

 activates above or below the set point

Hold the “SEL” key for approximately 6 seconds and “P-n1” will appear in the upper display, the control code will appear in the lower 
display, “0” or”1” will actuate 

below 

the set point, “2” or “3” will actuate 

above 

the set point. Press the “SEL” switch once, the value 

will flash, change the value with the “up” or “down “ keys, and press the “SEL” key again to set it in memory. Holding the “SEL” key 
for approximately 2 seconds will return to normal operation. 
5. 

How to change the Alarm Hysteresis (deadband ) 

Hold the “SEL” key for approximately 6 seconds and “P-n1” will appear in the upper display, Press "\/" down arrow and “A1hy” will 
appear, the hystersis amount will appear in the lower display, press the “SEL” key once and the amount will flash, change the 
amount to the desired value with the “up” or “down “ keys, press “SEL” again to set it in memory. Holding the “SEL” key for 
approximately 2 seconds will return to normal operation. 

 
 Calibration 

A. How to calibrate 

The front panel has two adjustments labeled "Zero" and "Span", always adjust the Zero first. Allow the controller to warm up for at 
least 10 minutes, Place the conductivity electrode in air, wait for the reading to stabilize and adjust the “Zero “ pot for a reading of 
00.00, place the electrode in a calibration solution of 10.00 uS, allow the reading to stabilize and adjust 
the “Span “ pot for a reading of 10.00.                                                                                                                                                               

Page2

 