QUICK START GUIDE
SENTIERO HANDHELD DIAGNOSTIC
(BLUE LABEL)
100800-06_Sentiero Handheld Diagnostic Quick Start Guide_Rev 2
Page
2
of
6
Select the appropriate ear tip size and place it on the probe tip. Insert the probe in the ear making
sure that the ear tip fits tightly into the ear canal.
Select the test ear by pressing either the Right or Left button. Cartoon mode will start the test showing
a cartoon instead of the progress bar.
Progress for
each tested
frequency
Overall test
progress
Noise floor
level
A calibration will be performed prior to measurement. A green light indicates successful
calibration (below left). After successful ear probe calibration, the test will start automatically.
Probe fit
override for
patients
with PE
tubes or
middle ear
perforations