Model No.PS-2205

 Setup

3

Xplorer and Xplorer GLX Setup

If you will be using an Xplorer or Xplorer GLX in logging mode (not connected to a 
computer), connect the Dual Load Cell Amplifier to the Xplorer or Xplorer GLX, turn 
the interface on, and press 

 to begin data collection.

SPARK Setup

If the SPARK Science Learning System (SLS) is off, press and hold the power 
button on the bottom to turn it on and then wait for the SPARK to boot up. The 
screen will show a message to plug in a sensor.

Connect the PASPORT sensor to either of the ports on the top of the SPARK. The 
screen will show the list of quantities measured by the connected sensor. 

Graph Display (default)

To open a graph display, touch any quantity in the list and then touch SHOW to open 
PAGE 1. Touch the right arrow next to PAGE 1 to go to the next display (digits). 
Touch the 

Start

 button to begin collecting data. 

Select a Display

To set up a particular display (e.g., digits display), touch BUILD. Touch a quantity 
from the list, and then touch one of the display icons. Touch OK to open the display, 
and then touch the 

Start

 button to begin collecting data.

Measurement Alignment

When you are using multiple Load Cells, you may find that their readings differ 
slightly when they are measuring the same thing. Before putting the Load Cells into a 
bridge, truss, or other structure, press the “TARE” button on the side of the Dual Load 
Cell Amplifier to “zero” the reading of each of the Load Cells.

No calibration should be necessary for the PASCO Load Cells connected to the Dual 
Load Cell Amplifier. The following description outlines the procedure for a one point 
simultaneous calibration using DataStudio.

Start button

See the User’s Guides for 
the Xplorer GLX or the 
SPARK Science Learning 
System for calibration 
instructions.

Содержание PS-2205

Страница 1: ...cted to the two input ports The Dual Load Cell Amplifier can be connected to any PASPORT interface One amplifier can be connected to a single USB Link or Xplorer Up to three amplifiers can be connected to a PowerLink and up to four amplifiers can be connected to the Xplorer GLX datalogger Included Equipment Part Number Dual Load Cell Amplifier PS 2205 Compatible Sensor Load Cell 100 N PS 2200 Load...

Страница 2: ... Dual Load Cell Amplifier in any order The maxi mum sampling rate is the same whether one or two Load Cell are connected Setup Hardware Setup The following three steps can be performed in any order 1 Connect one or two separate Load Cells PS 2200 to the individual input ports of the sensor box Load Cells can be connected to either port in any order The connector on the Load Cell cable fits into th...

Страница 3: ... 1 to go to the next display digits Touch the Start button to begin collecting data Select a Display To set up a particular display e g digits display touch BUILD Touch a quantity from the list and then touch one of the display icons Touch OK to open the display and then touch the Start button to begin collecting data Measurement Alignment When you are using multiple Load Cells you may find that t...

Страница 4: ...e for any load cell 9 96 mV Pin Signal 1 Strain Gauge Output 2 4 2 V 3 Strain Gauge Output 4 Ground 5 Range Select 1 6 Range Select 2 1 Click to open the Experiment Setup window Click to open the calibration dialog box 2 From the Sensor field select the Load Cell Amplifier 3 Select the Calibrate all similar measurements simulta neously option 4 Arrange all of the Load Cells so that the same amount...

Страница 5: ...nformation visit www pasco com legal Windows is a registered trademark of Microsoft Corporation in the United States and or other countries Mac is trademark of Apple Com puter Inc registered in the U S and other countries Patents Pending The following PASCO products have patents pending Product End of Life Disposal Instructions This electronic product is subject to disposal and recycling regulatio...

Отзывы: