Spirometer

Data Collection

4

®

Procedure Notes

Observing the following conditions will allow you to collect consistent and 
accurate data.

You may need to run the test a few times in order to get satisfactory 
data.

The subject should sit up or stand comfortably.

Talk the subject through each inspiration and expiration (this will 
prevent the subject from being distracted by counting breathes or 
remembering the sequence.)

Do not allow the test subject to see the data as it is collected.

Use a nose clip, or have the subject pinch his or her nose during the 
test.

Place the mouthpiece and handle on a table before you start data 
collection. Do not pick it up until the red WAIT indicator has stopped 
blinking and the green READY indicator is illuminated. While the 
WAIT indicator is blinking, the sensor is measuring the ambient 
pressure in order to establish a baseline reading. It is important not to 
expose the sensor to any pressure change or airflow during the wait 
period.

The electronics enclosure contains a high-resolution differential 
pressure sensing element that is sensitive to orientation. The orientation 
with respect to vertical or horizontal is not important, but it must 
remain constant during the wait period and data collecting.

Single-use mouthpieces are intended for a single test occasion, which 
may include several tests with the same individual. However, with each 
test, liquid and particles can accumulate on the screen of the 
mouthpiece, which increase the resistance to air flow, leading to 
unrealistically high readings of Flow Rate and Total Volume in 
subsequent tests. Replace the mouthpiece if you observe evidence of 
screen contamination. For the most accurate results, use a new 
mouthpiece for every test.

Содержание PASPORT PS-2152

Страница 1: ...ory volume and other important lung volumes and capacities Included Equipment Part Number Electronics enclosure and handle PS 2152 Single use mouthpiece 2 included PS 2522 10 pack CD ROM containing Da...

Страница 2: ...collect and display data for a pulmonary function test On screen instructions will guide students through data collection and analysis Students can review the test procedure and interpretation of PFT...

Страница 3: ...green READY indicator has illuminated place the tapered end of the mouthpiece between the test subjects s front teeth with the lips creating a seal to ensure that all inspired and expired air flows th...

Страница 4: ...T indicator is blinking the sensor is measuring the ambient pressure in order to establish a baseline reading It is important not to expose the sensor to any pressure change or airflow during the wait...

Страница 5: ...level of tidal expiration Inspiratory Reserve Volume IRV is the amount of air moved into the lungs during a forced inspiration beyond the level of tidal inspiration Forced Expiratory Volume in One Sec...

Страница 6: ...forced expiration A typical RV is 1 2 L Functional Residual Capacity FRC is the amount of air left in the lungs after a normal expiration FRC RV ERV A typical FRC is 2 5 liters The formula for Lung V...

Страница 7: ...apacity measurements taken from the graph since each measurement is made between two data points recorded within a few seconds In DataStudio you can eliminate the trend with this calculation Total Flo...

Страница 8: ...tance with any PASCO product contact PASCO at Address PASCO scientific 10101 Foothills Blvd Roseville CA 95747 7100 Phone 916 786 3800 800 772 8700 Fax 916 786 3292 Web www pasco com Email techsupp pa...

Отзывы: