background image

 

 
 
 

Wet Bath Accuracy Checking

 

Using  wet  bath  simulators  for  accuracy  checking  has  been  the  accepted  method  for 
many  years.    Breath  alcohol  simulators  are  specially  designed  water-alcohol 
instruments which provide equilibration of alcohol between water and air at a controlled 
temperature. 
 
Accuracy  checking  of  the  MARK  V  should  be  performed  by  authorized  persons  using 
any NHTSA approved breath alcohol simulator. 
 

To perform a wet bath accuracy check: 

1.  Pour 500 ml of 0.080% Certified Solution into the glass jar. 
2.  Attach a piece of tubing (6 - 

8” long) to simulator inlet.  Attach a regular or check-

valve mouthpiece to the end of this tubing. 

3.  Attach a 1 - 

2” piece of tubing to the simulator outlet.  Attach a mouthpiece to the 

other end of this shorter tubing.  Make sure the connection is air-tight. 

 

Note:    The  length  of  tubing  connected  to  the  simulator  outlet  should  not  be 
longer than 2” in order to prevent condensation. 

 

4.   Plug Simulator in and turn switch to the ON position. 
5.  Allow the solution to heat. 
6.  After  15  -  20  minutes,  check  the  thermometer.    The  thermometer  should  read 

34ºC when ready. 

7.  Attach  instrument  to  the  mouthpiece.    (The  opaque  moisture  trap  can  be  used 

should condensation appear in the mouthpiece). 

8.  Switch on the MARK V.  Wait for 

“Please Blow” to display.  Connect to outlet tube. 

9.  Forcefully blow into the simulator for several seconds, press the green button and 

release. 

 

After  a  few  seconds,  the  measured  value  will  be  displayed.    The  display  should  read 
0.080% +/-0.005. 

 

NOTE:    If  the  result  is  not  within  published  specifications  for  the  MARK  V  ,  conduct 
another  accuracy  check.    Verify  that  the  seal  on  the  simulator  is  air-tight  and  that  the 
outlet  tube  and  mouthpiece  are  free  of  condensation.    If  the  measurement  is  still  not 
within specifications (+/- 0.005), the unit must be removed from service until an internal 
calibration can be conducted. 
 

 

 
 

 

Содержание MARK V Alcovisor

Страница 1: ...K V Alcovisor Operating Manual PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 www pasintl com 800 660 SNIF 540 372 3431 Copyright 2011 by PAS Systems International Inc A...

Страница 2: ......

Страница 3: ...or Test Subject 5 Air Blank Feature 5 Measuring BrAC BAC 6 Automatic Sampling 6 Manual Sampling 7 Conducting a Passive Test No Mouthpiece 7 Accessing Test Records 8 Accuracy Checking and Calibration 8...

Страница 4: ......

Страница 5: ...cedure to be conducted in about 30 seconds or less This manual describes the operation maintenance calibration check and calibration adjustment of the MARK V This manual should be read completely and...

Страница 6: ...r and displayed on screen The instrument is simple to operate and may be used as often as required provided that a suitable delay is allowed between successive tests This time delay allows the fuel ce...

Страница 7: ...e cover When replacing the batteries always replace ALL four batteries Turning On Off ON Press and hold the Green ON OFF button for 2 3 seconds The display light comes on and a self test automatic Bla...

Страница 8: ...rea and RIGHT green buttons to select the correct Date and Time b Select Update to confirm 2 Auto Off a Use the LEFT red and RIGHT green buttons to Select the time limit to turn off the device automat...

Страница 9: ...e test The person must also avoid repeated deep breaths hyperventilation as this will temporarily cool down the breath and may lead to a false reading Should the person have diminished lung capacity a...

Страница 10: ...click sounds Blowing can cease when the click is heard 7 After a few seconds the Test Result will be displayed on the LCD If the unit is left on it will power down after displaying the test result fo...

Страница 11: ...RK V ON 2 When Please Blow appears hold the sample port mouthpiece opening about 4 inches from the subject s mouth 3 Have the subject moderately blow towards the mouthpiece holder opening for several...

Страница 12: ...est is not within the acceptable range the unit must be re calibrated by PAS Systems International Inc or by persons who have been properly trained for alcohol detector calibration Due to the sensitiv...

Страница 13: ...nnected to the simulator outlet should not be longer than 2 in order to prevent condensation 4 Plug Simulator in and turn switch to the ON position 5 Allow the solution to heat 6 After 15 20 minutes c...

Страница 14: ...red value will be displayed If the measured value is within 0 005 BrAC of the altitude adjusted concentration value the MARK V is working accurately Note Suitable Dry Gas Standards and Regulators are...

Страница 15: ...0 080 Dry Gas 0 1 0 080 500 0 981 0 078 1000 0 962 0 077 1500 0 943 0 075 2000 0 925 0 074 2500 0 907 0 073 3000 0 889 0 071 3500 0 872 0 070 4000 0 854 0 068 4500 0 837 0 067 5000 0 820 0 066 5500 0...

Страница 16: ...There is no limitation on the number of tests that may be conducted between calibration checks providing the monthly checks are completed 3 External Calibration Check Tolerance 0 005 or 5 at or above...

Страница 17: ...call PAS Systems International Inc technical service at 800 660 7643 PROBLEM CAUSE SOLUTION The display is only faintly lit or the unit will not turn on The power supply to the unit is completely exh...

Страница 18: ...Temperature 23 to 131 F 5 to 55 C Storage Conditions 13 to 158 F 25 to 70 C not more than 95 relative humidity LCD Display Size 2 0 X 1 5 viewing area 5cm X 4 cm Automatic backlight for night time us...

Страница 19: ...onal Inc or an authorized service technician Only original MARK V parts may be used To keep the instrument clean periodically use a mild disinfectant and a soft cloth on the outside of the case DO NOT...

Страница 20: ...MARK V Alcovisor Law Enforcement PAS Systems International Inc 215 Southport Dr Suite 400 Morrisville NC 27560 http www pasintl com Tel 540 372 3431 800 660 SNIF Printed 10 12 5 00...

Отзывы: