MA-1040 Magnetic Analyzer

Cleaning the Sample Holder 

May 2011                                                 

7

Cleaning the Sample Holder

Sample holders should always be clean and dry before adding sample. Use a mild soap and water to  
clean the sample holder; a soft-bristled brush may also be used. 

Weighing the Sample

Perform the following steps to obtain the sample weight:

1. Place a clean, dry sample holder on the balance. Allow the balance to stabilize then tare it to zero.

2. Using a proper lab technique (such as stirring), mix the sample to evenly distribute all grains.

3. Remove the sample holder from the balance and fill with sample until it is just below the over-fill 

orifice.

4. Place the sample holder (containing sample) back onto the balance. Allow the balance to stabilize 

and record the weight shown on the display.

Do not use steel brushes or harsh solvents (such as acetone) to clean the 
sample holder. Doing so will scratch the inside of the sample holder.

Be sure to mix the sample well so that grains are distributed evenly. Settling of 
the sample will result in an inaccurate reading. 

Содержание MA-1040

Страница 1: ...MA 1040 Magnetic Analyzer Operator s Manual V2 0 P03 42801 00 March 2011...

Страница 2: ...Particulate Systems 2011 All rights reserved Printed in the U S A...

Страница 3: ...products sold by MICROMERITICS are the products of reputable manufacturers sold under their respective brand names or trade names We therefore make no express or implied warranty as to such products...

Страница 4: ......

Страница 5: ...f Contents Introduction 1 Description 1 Components and Connectors 2 Top Panel 2 Side Panel 3 Specifications 4 Setting up the Analyzer 5 Reference Standard Test 6 Cleaning the Sample Holder 7 Weighing...

Страница 6: ......

Страница 7: ...g insula tion The MA 1040 can also be used to detect low magnetic iron levels in food precious gems pharmaceuticals and many other materials The MA 1040 has been referenced in American National Standa...

Страница 8: ...s of magnetic content in PPM WEIGHT GRAMS control Used to enter the sample weight Range 000 999 grams ZERO control Used to zero the reading on the display when the sample chamber is empty Sample Chamb...

Страница 9: ...MA 1040 Magnetic Analyzer Description May 2011 3 Side Panel Power connector For connecting the power cord On Off switch For turning the analyzer on and off Power connector On Off switch...

Страница 10: ...rature Stable between 10 and 45 C 50 to 113 F operating 10 and 55 C 14 to 131 F non operating Physical Height 16 76 cm 6 6 in Width 25 40 cm 10 in Depth 13 34 cm 5 25 in Weight 6 8 kg 15 lbs Electrica...

Страница 11: ...r on a steel table If unavoidable place a 3 4 in thick piece of plywood underneath the instrument Ensure a constant room temperature fluctuations may result in inaccurate readings 1 Place the analyzer...

Страница 12: ...ote of each reading 5 Take an average of the three readings from steps 3 and 4 The average for the three readings should be within 1 of the value printed on the reference standard or 0 5 PPM whichever...

Страница 13: ...o zero 2 Using a proper lab technique such as stirring mix the sample to evenly distribute all grains 3 Remove the sample holder from the balance and fill with sample until it is just below the over f...

Страница 14: ...rate for about three to five seconds after which time the display may drift 4 Remove and empty the sample holder Place the empty sample holder back into the sample chamber and record the reading shown...

Страница 15: ...rder Part Number Item and Description P03 25803 00 Sample holder P03 34810 0010 Reference standard 10 PPM P03 34810 0050 Reference standard 50 PPM P03 34810 0100 Reference standard 100 PPM P03 34810 0...

Страница 16: ......