Installation
Chapter 2 Unpacking and Installation
MiniCapt Remote Microbial Air Sampler Operations Manual
Page 2-5
•
Environmental conditions
The unit should be located in an environment consistent with its temperature,
humidity, altitude, and other relevant specifications.
•
Electromagnetic interference
The unit should be located away from strong electromagnetic fields.
Positioning or Mounting the Unit
The MiniCapt Remote Microbial Air Sampler can be mounted to a flat surface with a
mounting bracket that attaches to the bottom of the sensor’s chassis with
machine screws provided with the mounting bracket (see
).
CAUTION
The mounting location MUST be sufficiently strong.
Do NOT mount the instrument to unreinforced sheetrock.
Power
Exhaust Port
Compressed
Gas Ports
Connector
Ethernet
Connector
Mounting Bracket
Serial/Service Port
(remove the screws)
Figure 2-1
MiniCapt Remote Microbial Air Sampler attached to a mounting bracket
NOTE:
The main power disconnect is the power cord, so ensure that the
power cord plug is easily accessible.