BioCapt

®

 Adjustable Height Quick Guide

1. 

Remove sampler head.

2. 

Turn plate holder counterclockwise until flush with rim  

(2-3 clicks from fully loose).

3. 

Turn plate holder clockwise for:

• 

4 complete turns (32 ml pour)

• 

3 complete turns (25 ml pour)

• 

2 1/2 turns (18 ml pour)

• 

Finely adjust with equipped agar plate if needed. 

Lower 5 clicks after agar contact with impactor head.

P L A T E   H E I G H T

1. 

Remove sampler head and plate holder.

2. 

Lay the plate holder upside down to see screws. 

3. 

Use 1.5 mm hex driver to:

• 

Loosen screws for fine adjustment of bracket 

OR

• 

Remove screws to align bracket to other set of holes.

4. 

Flip plate holder upright. 

5. 

Center agar plate inside brackets using etched circles.

6. 

Tighten all screws with 1.5 mm hex driver.

7. 

Reinstall plate holder and sampler head.

P L A T E   W I D T H

TA

LL

ER

 PL

ATE

PLATE HOLDER

S H O R T

E R

 P

LA

TE

SCREWS

BRACKET

OUTER HOLES

INNER HOLES

ETCHED CIRCLES