Incubate for at least 30 minutes. Re-start the
system by pressing START and let it run to the
end. The Decontamination Solution may incubate
even over night! Wash the flow cuvette by
attaching a tube with 1.6 ml of Partec Sheath
Fluid and run it for at least 2 minutes.
In case your flow cytometer shows degraded
performance even after thorough cleaning, please
contact your local distributor or Partec.
Maintenance and Service
Maintenance
Clean the CyFlow
®
space casing on a regular base, carefully with
soft cloth. Water must not enter the CyFlow
®
space or peripheral
devices or come into contact with electric connections and
switches. For cleaning the screen, always use special screen
cleaner and soft cloth.
Do not use
any organic solvents, nitro thinner, benzol, alcohol,
highly concentrated bleach etc!
For cleaning of flow cuvette, refer to page 21. Do not use tools to
clean the flow cuvette. In case the flow cuvette is blocked,
enquire Partec for rapid exchange.
Regularly empty the waste bottle and clean with warm detergent
solution and a brush.
Clean sheath reservoir with destilled water and a clean brush and
flush with clean destilled water several times. Remember
cleanliness of sheath fluid reservoir is critical for proper
operation.
If the CyFlow
®
space will not be used for longer periods, clean
flow system by using distilled water. Put a sample tube half-ways
filled with destilled water at the sample port. Clean waste and
sheath reservoir, wipe top dry.
Service
All service is to be made from an authorized service engineer.
Please contact your supplier or partec for service requests.
Transport and Storage
For the transport of the system to a different
location it will be necessary to disconnect all
external tubing, data, and supply connections. In
case of use with potentially biohazardous
material, please see Partec standard operating
procedure (SOP) for decontamination. The
system should be carried in upright position.
During transport or storage please take care that
the system will be stored under the following
conditions:
Temperature
5-50 C
Humidity
20-85% relative (non-condensing)
Room
Clean environment, no direct sun
light
Disposal
In case of product disposal, please proceed
according to the Partec standard operating
procedure (SOP) for decontamination.
After decontamination, the system has to be
disposed according to the local regulations and
laws.
For further information, please contact your local
distributor or Partec.