K3.1.114o Manual NitroFlow Mobile
- 24 -
SCREEN 3/5
Data Explanation
Outlet
Outlet pressure in either BAR or PSI
Inlet
Inlet/compressor pressure in either BAR or PSI
C Inlet
Compressed air inlet temperature in °C or °F
SCREEN 4/5
Data Explanation
Remote
YES or NO
Shows whether remote control option is on or off
SCREEN 5/5
Button Explanation
O2 – 20.9%
Calibrate O
2
sensor to 20.9%. Contact Parker’s technical support team.
Flow Factor
Only visible when selected Show Flow in the options menu
(refer to §6.3.4)
and when the unit is running. Calibrate the flow by entering the flow measured
with an external flow meter.
Replace Filter
When a filter has been replaced during maintenance, this button can be touched
and the countdown for the new filter is set. System asks for confirmation. In
maintenance screen 1/5 the filter lifetime should read 8000 hr.
Continues on next page.