background image

Parker Hannifin Corporation  

Industrial Gas Filtration and Generation Division  
Haverhill, MA 

•  1-800-343-4048

 

Parker Zero Air Generators 

Models HPZA-3500, HPZA-7000, HPZA-18000, HPZA-30000

 

 

Bulletin TI-HPZA-3500H

 

 

     

 

Optional Accessories

 

 

Model Number 

Description 

 

W-405-4032-000 

Pressure Regulator 

 

 

W-FM Series 

Flow Controllers 

 

76080 

Halogenated Hydrocarbon Scrubber 

 

 

 

 

System Specifications

 

HPZA-3500

 

HPZA-7000

 

HPZA-18000

 

HPZA-30000

 

Max flow rate

 

3.5 lpm

 

7 lpm

 

18 lpm

 

30 lpm

 

CSA Certification Standard

 

         CAN/CSA 22.2 No. 61010-1-12 & 61010-2-010

 

IEC Safety Standard 

IEC 61010-1:  2010 + A1 2019 + A2: 2010 / EN 61010-1:  2010  

 

UL Safety Standard 

UL 61010-1 & 61010-2-010 

 

ETL listed to U.S. and Canadian Safety Standards

 

Total hydrocarbon content

 

<0.05 ppm

 

<0.05 ppm

 

<0.05 ppm

 

<0.1 ppm

 

Min/Max inlet pressure

 

40 psig/125 psig

 

2.8 barg / 8.6 barg

 

40 psig/125 psig

 

2.8 barg / 8.6 barg

 

40 psig/125 psig

 

2.8 barg / 8.6 barg

 

40 psig/125 psig

 

2.8 barg / 8.6 barg

 

Min/Max ambient temperature

 

60°F (16°C) / 100°F (38°C)

 

60°F (16°C) / 100°F (38°C)

 

60°F (16°C) / 100°F (38°C)

 

60°F (16°C) / 100°F (38°C)

 

Max relative humidity

 

80%

 

80%

 

80%

 

80%

 

Max inlet hydrocarbon content

 

100 ppm

 

100 ppm

 

100 ppm

 

100 ppm

 

Pressure drop at max flow rate

 

4 psid (0.3 bar)

 

4 psid (0.3 bar)

 

4 psid (0.3 bar)

 

4 psid (0.3 bar)

 

Start-up time

 

45 min.

 

45 min.

 

45 min.

 

45 min.

 

Electrical requirements 

120 VAC, 220 VAC, 230 VAC, 240 VAC, ±10% nominal,  50/60 Hz models available.

 

See product label on generator for specifications.

 

 

Current rating 

1.0 A at 220 VAC, 2.0 A at 120 VAC 

2.0 A at 220 VAC, 4.0 A at 120 VAC

 

Shipping weight

 

41 lbs. (19 kg)

 

41 lbs. (19 kg)

 

41 lbs. (19 kg)

 

41 lbs. (19 kg)

 

Physical dimensions

 

11”w

 x 

13”d

 x 

16”h

 

27 cm x 34 cm x 42 cm

 

11”w

 x 

13”d

 x 

16”h

 

27 cm x 34 cm x 42 cm

 

11”w

 x 

13”d

 x 

16”h

 

27 cm x 34 cm x 42 cm

 

11”w

 x 

13”d

 x 

16”h

 

27 cm x 34 cm x 42 cm

 

Altitude

 

2000 m

 

2000 m

 

2000 m

 

2000 m

 

Fuse Type 

Type T, 250 V, 1.0 A at 220 VAC, 2.0 A at 120 VAC 

Type T, 250 V, 2.0 A at 220 VAC, 4.0 A at 120 VAC

 

Содержание HPZA-18000

Страница 1: ...e 1 are completely engineered systems which replace the use of inconvenient high pressure gas cylinders as a source of hydrocarbon free air The HPZA 30000 generators purify on site compressed air to a...

Страница 2: ...rs are equipped with float drains which automatically open to empty any liquids from inside the filter housing The drains are connected to 1 4 O D plastic tubing which discharges to atmosphere onthe r...

Страница 3: ...re taken into consideration Approximately 6 of space should be allowed along the top and sides of the generator to ensure proper ventilation Compressed Air The Parker Zero Air Generator requires a sou...

Страница 4: ...ion only Piping Components The inlet and outlet ports for the Parker Zero Air Generator are 1 4 NPT female and located on right side of the unit Inlet tubing and fittings should be clean and rated for...

Страница 5: ...stream from the generator Exposure of the generator to halogenated hydrocarbons will damage the catalyst module and void the warranty For additional information request the Analytical Gas Supplies cat...

Страница 6: ...carbon content of less than 0 05 ppm Models HPZA 3500 HPZA 7000 and HPZA 18000 a flow control device must be installed downstream from the generator to prevent downstream demand in excess of the gener...

Страница 7: ...Replace the GS membrane cartridge GS 050 05 95 and re assemble The time required to service all three stages of filtration integral to the Parker Zero Air Generator is approximately 15 minutes To prev...

Страница 8: ...tor may burn out The fuses are located in the power receptacle on the left hand side of the generator Before servicing the fuses turn the generator off and disconnect the power cord from both the powe...

Страница 9: ...generator is exceeded Overflow LED and yellow THC LEDs will illuminate An excessive pressure drop greater than 4 psig 0 3 bar may accompany an overflow situation If demand exceeds capacity install a f...

Страница 10: ...arg 40 psig 125 psig 2 8 barg 8 6 barg 40 psig 125 psig 2 8 barg 8 6 barg Min Max ambient temperature 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C Max relative h...

Страница 11: ...calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or i...

Страница 12: ...800 343 4048 Symbol Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 242 Neck Road Haverhill MA 01835 www parker com labgas Description Caution Refer to accompanying docum...