Parker H2PEMPD-510 Скачать руководство пользователя страница 5

Bulletin TI-H2PEMPD-D 

 

Installation, Operation and Maintenance Manual 

 

 

Hydrogen Gas Generator H2 PEMPD series 

 

1-800-343-4048 

 

www.parker.com/balston 

1 Safety Information  

 

 

 

 

 

 

 

 

         

Do not install and operate this equipment until the safety information and instructions in this user guide have been read and understood by 
all personnel concerned. 

 

USER RESPONSIBILITY  

FAILURE OR IMPROPER SELECTION OR IMPROPER USE OF THE PRODUCTS DESCRIBED HEREIN OR RELATED ITEMS CAN CAUSE DEATH, 
PERSONAL INJURY AND PROPERTY DAMAGE.  

This document and other information from Parker Hannifin Corporation, its subsidiaries and authorized distributors provide product or system 
options for further investigation by users having technical expertise.    

The user, through its own analysis and testing, is solely responsible for making the final selection of the system and components and assuring 
that all performance, endurance, maintenance, safety and warning requirements of the application are met. The user must analyze all aspects 
of the application, follow applicable industry standards, and follow the information concerning the product in the current product catalogue 
and in any other materials provided from Parker or its subsidiaries or authorized distributors.    

To the extent that Parker or its subsidiaries or authorized distributors provide component or system options based upon data or specifications 
provided by the user, the user is responsible for determining that such data and specifications are suitable and sufficient for all applications and 
reasonably foreseeable uses of the components or systems.  

Only competent personnel trained, qualified, and approved by Parker Hannifin should perform installation, start-up, service and repair 
procedures.  

This equipment is for indoor use only. Do not operate outdoors.  

This equipment is not suitable for use in any Hazardous, Flammable, or Explosive environments. Hydrogen is a highly flammable gas. Keep the 
generator away from excessive heat and naked flames.  

With the exception of oxygen, any gas can cause asphyxiation in high enough concentrations. In most scenarios, however, because hydrogen 
rises and disperses so rapidly, it is unlikely to be confined where asphyxiation might otherwise occur. Always ensure that the generator is 
operated in a well-ventilated area and all of the vent ports on the rear of the generator are kept clear and free from blockages.  

Use of the equipment in a manner not specified within this user guide may result in an unplanned release of pressure, which may cause serious 
personal injury or damage.  

When handling, installing or operating this equipment, personnel must employ safe engineering practices and observe all related regulations, 
health & safety procedures, and legal requirements for safety.  

Ensure that the equipment is depressurized and electrically isolated, prior to carrying out any of the scheduled maintenance instructions 
specified within this user guide.  

Parker Hannifin cannot anticipate every possible circumstance which may represent a potential hazard. The warnings in this manual cover the 
most known potential hazards, but by definition cannot be all-inclusive. If the user employs an operating procedure, item of equipment or a 
method of working which is not specifically recommended by Parker Hannifin the user must ensure that the equipment will not be damaged or 
become hazardous to persons or property.  

Most accidents that occur during the operation and maintenance of machinery are the result of failure to observe basic safety rules and 
procedures. Accidents can be avoided by recognizing that any machinery is potentially hazardous.  

Note: Any interference with the calibration warning labels will invalidate the gas generator's warranty and may incur costs for the recalibration 
of the gas generator.  

Should you require an extended warranty, tailored service contracts or training on this equipment, or any other equipment within the Parker 
Hannifin range, please contact your local Parker Hannifin office.  

Details of your nearest Parker Hannifin sales office can be found at www.Labgasgenerators.com   

Retain this user guide for future reference.

 

Содержание H2PEMPD-510

Страница 1: ...enance Manual Hydrogen Gas Generator H2PEMPD Series www parker com fns balstonlabgenerators 1 1 800 343 4048 Parker Hydrogen Gas Generator Installation Operation and Maintenance Manual Models H2PEMPD...

Страница 2: ...warranty period of 1 year from the date of shipment WARNING The H2PEMPD 1300 hydrogen generator operating at full flow will consume the entire water reservoir volume in approximately 60 hours If the...

Страница 3: ...ision of a replacement part provided that the equipment has been used strictly in accordance with the instructions provided with each item of equipment and has been stored installed commissioned opera...

Страница 4: ...y generators fitted with Auto Water Fill 9 3 3 Options Board Accessory 9 3 3 1 Fitting the Options Board 9 3 3 2 Wiring the Options Board 10 3 4 Starting up the Generator 11 4 Operating the equipment...

Страница 5: ...and naked flames With the exception of oxygen any gas can cause asphyxiation in high enough concentrations In most scenarios however because hydrogen rises and disperses so rapidly it is unlikely to...

Страница 6: ...ration and Maintenance Manual Hydrogen Gas Generator H2 PEMPD series 2 1 800 343 4048 www parker com balston 1 1 Markings and Symbols The following markings and international symbols are used on the e...

Страница 7: ...l environments and is nonhazardous for transportation purposes In order to guarantee the optimum efficiency of the PEM cell this generator must be installed and running within three months of dispatch...

Страница 8: ...Category Maximum Altitude Noise Supply Voltage Range Frequency Fuse 4 Environmental Data Ambient Temperature Relative Humidity IP Rating Deionized ASTM II 1M 1 S Filtered to 100 m 1 45 0 1 1000 68 20...

Страница 9: ...Note The storage area should be secure and the environmental conditions should fall within those specified in the technical specification If the generator is stored in an area where the environmental...

Страница 10: ...in 2 Options board connection port 9 Fused IEC 320 inlet socket and ON OFF switch 3 O2 vent 250ml min 10 Water bottle spillage drain 4 Excess H2 vent 1 ml min 11 Environmental filter 5 Water bottle ve...

Страница 11: ...an displace oxygen thereby creating an asphyxiation hazard Always ensure that the equipment is operated in a well ventilated area 2 4 4 Water Supply Requirements Generators fitted with an automatic wa...

Страница 12: ...e procedures 3 1 Recommended system layout Figure 3 Table 6 Note For applications that require multiple generators please contact your local Parker Hannifin office 3 1 1 Installation Kit Kit Number De...

Страница 13: ...ist When routing the tubes ensure that they are adequately supported to prevent damage and leaks in the system All components used within the system must be rated to at least the maximum operating pre...

Страница 14: ...with Auto Water Fill The optional water fill allows the generator s water bottle to be gravity fed from a suitable deionized water supply When the water level falls below the mid point the water bott...

Страница 15: ...to an external stop circuit Press to reset the generator JP4 Alarm Output The alarm output is designed for remote alarm indication When an error occurs on the generator the output switching circuit i...

Страница 16: ...completion the generator will warm up and the display will show Initializing Once warm up is complete the unit will revert to the default menu as shown in figure 10 Note Upon first use the generator...

Страница 17: ...is happen restart the generator The internal pressure ACT pressure of the generator will build up to the required operating pressure SET pressure Once the required pressure is reached the outlet valve...

Страница 18: ...essure of the generator LOAD X Indicates the status of the generator outlet valve X Outlet is closed Outlet is open 100 The shaded blocks indicate the rate of hydrogen production Each block represents...

Страница 19: ...drops to four the Change Water error message will be displayed the water bottle indicator will flash yellow and an intermittent alarm will sound Hydrogen will continue to be delivered to the applicat...

Страница 20: ...in liters Figure 15 4 3 6 Error Log The Error Log menu allows the user to access the 10 most recent error messages To access the errors press and hold The menu will display the most recent error messa...

Страница 21: ...uidance When the faults are rectified switch the generator off at the mains switch Press and hold At the same time reapply the power to the generator When the generator has powered up press again The...

Страница 22: ...on the equipment 5 1 1 Service Intervals Table 11 Component Daily 6 months 4000 Hrs 24 Months 16000 Hrs Generator Generator Generator Generator Generator Generator Generator Generator Service 6 month...

Страница 23: ...able 12 5 2 2 Recommended Service B Required every 16000Hrs 24 months Table 13 Description Kit No Contents 1 Deionizer cartridge 2 100 micron water filter 3 Environmental filters x2 4 Filter replaceme...

Страница 24: ...cap Wearing disposable gloves remove the deionizer cartridge figure 18 item 1 and discard Extract the 100 micron water filter figure 18 item 2 using the H2 filter replacement tool figure 18 item 4 Pu...

Страница 25: ...re 22 Remove the outer heater box being careful of the internal tubing connection Slide the top rear cover off and store the panel safely as shown in figure 22 With the Environmental filters still rem...

Страница 26: ...ered to 100 m within this generator will damage and reduce the life time of the hydrogen cell Remove the top front cover and the water bottle cap Fill the water bottle using fresh deionized water to a...

Страница 27: ...rain and refill with Deionized ASTM II 1M 1 S filtered to 100 m Cell voltage HIGH Off Reset or hard reset Contact tech support for cell or PSU replacement if the problem persists Cell voltage LOW Off...

Страница 28: ...out of date On Calibrate conductivity sensor Software timeout Off Reset contact tech support Pump fault no cell water circulation Off Check replace pump or connection Power supply fan not rotating Of...

Страница 29: ...eater Purifier temp HIGH Off Reset check connector Replace heater Purifier temp LOW Off Reset check connector Replace heater Service time On Perform service Use applicable kit O2 vent is blocked Off S...

Страница 30: ...482 6296 www parker com pag dhFNS_TSG parker com Parker Hannifin Corporation Filtration and Separation Division 242 Neck Road Haverhill MA 01835 Tel 978 858 0505 Fax 978 478 2501 www parker com balsto...

Отзывы: