switching power supply adapter with Input being 100-240 VAC, 
1.0-2.0 Amp, 50/60 Hz and Output being 24 VDC, 3.0 Amp.

•  Insert the 2.5 mm metal plug from the Power Adapter into the 

back of the scaler. 

•  Insert the supplied AC cord female plug into the Power Adapter.
•  Plug the male AC power cord into any grounded electrical outlet.

Use of 30KHz 

Detachable Handpiece

The TurboSensor® + is provided with a 30KHz detachable handpiece. 
The handpieces do not come sterilized and must be autoclaved 
prior to initial use.

To attach the cable to the handpiece, align the small groove in the plug 
with the mating groove in the jack and push the connector gently until 
it “clicks”. Removal is accomplished by rotating the hose collar and 
gently withdrawing the connector from the handpiece.

Turning on the Scaler and 

Controlling the Power Output

The unit provides a rocker switch on the rear panel for Main Power 
Control. UP position of the switch turns the device ON. Green LED 
on the front panel would illuminate indicating powered ON status 
of the device. DOWN position of the switch on the rear panel would 
turn the device and Green LED indicator OFF.

Scaling will be activated upon pressing the foot switch once the 
device is turned ON. The pedal should be placed where it will not 
accidentally be pressed when the device is not in use. 

The Turb gives you two ways to adjust the scaling power, 
by using the Power Control Knob for normal operation, or by 
depressing the Foot Pedal to enter the “Turbo Mode”. 

POWER

Use the Power Control Knob to set the initial 
scaling power. Rotate the knob in a 
clockwise direction to increase the power, 
or counterclockwise to decrease the power.

When starting a procedure, use the Power 
Control Knob to set the power to “Medium”. 
When the foot pedal is slightly depressed to 
the fi rst position, the insert will begin 

vibrating at medium power. Then adjust the power up or down based 
the specifi c needs for the case being treated

Use the “Turbo Mode” for short-term increases in power during 
scaling. Using increased foot pressure, fully depress the foot pedal 

to the fl oor to enter “Turbo” mode.  A yellow indicator light will 
illuminate on the front panel to confi rm the scaler is operating at 
“Turbo” power. Scaling power to a point midway between the 
current setting and the scaler’s maximum power.

Engaging the Turbo feature at a low power setting causes a 
signifi cant boost in power.

Engaging the Turbo feature at high power will produce very 
little change.

Since water fl ow is not aff ected by “Turbo Mode”, extended 
procedures may require a water adjustment.

Partially lifting your foot off  the pedal will return the scaler to 
normal, “non- Turbo” mode. Releasing your foot completely off  the 
pedal will deactivate scaling. Green LED indicator will NOT turn 
OFF, as it indicates device ON status.

Purging

Per CDC recommendations, waterlines should be purged based on 
the following schedules:

•  Two minutes at the start and end of each day.
•  20 seconds between patients.

The Turb off ers the following 

exclusive Purge features:

•  Purging ONLY with the handpiece detached from the cable to 

prevent contamination of an autoclaved handpiece.

•  Convenience of a foot controlled “Tap-On” Purge.

How to Purge

1. 

Confi rm the unit is ON and that the green LED on the front 
panel is illuminated.

2. 

Disconnect the detachable-autoclavable handpiece from 
the cable.

3. 

Adjust the water control to Maximum fl ow.

4. 

Hold the cable end over a sink or drain.

5. 

Depress the foot pedal for more than 2 seconds to activate 
purge mode.
a. 

When initial pressure is applied to the foot pedal, a short 
burst of water will be expelled from the cable to confi rm 
purge mode is being activated.

b. 

After two seconds, a steady stream of water will fl ow, and 
the Purge LED will begin to pulse.

c. 

Once water fl ow is continuous, and the Purge indicator 
LED pulses, foot pedal pressure can be released without 
interruption of the purge cycle.

6. 

The Purge light will pulse for the fi rst 20 seconds of the purge 
cycle after which it will stay constantly on for an additional 1 
minute 40 seconds. (2 minutes of total purge time).

7. 

When the Purge light changes from pulse to constant the 20 
second purge cycle for between patient waterline purging is 
complete.  Depress the foot pedal to stop purging.

8. 

The beginning and end of day purge cycle will automatically 
shut off  after 2 minutes.

Safety Feature: 

 Deactivation of the purge cycle can be manually 

accomplished at any time by tapping the foot pedal.

Содержание TurboSensor +

Страница 1: ...idered a substitute for formal training in ultrasonic scaling Appropriate dental professional education is strongly recommended prior to using this device clinically A01156revA0722 English Instructions for Use Parkell Technical Support Service 800 243 7446 M F from 8 30 AM to 5 00 PM EST ...

Страница 2: ... of Power Cord 110V 12 6 380cm 220 230V 7 6 230cm Handpiece Operating Frequency 30 KHz hD722 Protection Against Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes on 5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storag...

Страница 3: ... handpiece cable and the handpiece are clean and dry before assembling them for use As with all electrical devices the unit should not be immersed in fluids and fluids should not be allowed to enter the unit through front or rear panel slots Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device Return it to Parkell for servicing Modifying de...

Страница 4: ...foot from pedal to evaluate deposit removal Exercise caution near ceramic or resin restorations as they can fracture if stressed Inserts Magnetostrictive Inserts for the TurboSensor scaler are available separately and are not included with the basic scaler unit The unit is designed to work with all Parkell and other market leading 30KHz inserts Brand new inserts may fit very tightly in the handpie...

Страница 5: ...o enter Turbo mode A yellow indicator light will illuminate on the front panel to confirm the scaler is operating at Turbo power Scaling power to a point midway between the current setting and the scaler s maximum power Engaging the Turbo feature at a low power setting causes a significant boost in power Engaging the Turbo feature at high power will produce very little change Since water flow is n...

Страница 6: ...ace You may gouge the tooth Pause occasionally during scaling by removing foot from pedal to evaluate deposit removal Exercise caution near ceramic restorations as they can discolor or fracture if stressed Common Operator Errors that May Result in Poor Clinical Performance Failure to keep insert tip parallel to the long axis of the tooth Excessive use of hand pressure Failure to use the lowest eff...

Страница 7: ...logged Inspect and clear blockage or replace insert Water filter clogged Replace No water supply connected Correct Insert does not go into handpiece Insert stack is bent Straighten plates carefully by hand and reinsert O ring is too tight Lubricate O ring with water and use twisting motion to seat insert Detachable 30K Handpiece hD722 can only accept 30K insert Make sure that it is not a 25KHz ins...

Страница 8: ...essories with the unit and include an explanation of the problem Pack in the original box add plenty of cushioning material and over box the unit during shipping Transit should maintain a dry temperature of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must be ma...

Отзывы: