background image

Trio SCF IFU Rev E 3/31/10 2:05 PM Page 11 

Composite

C

M

Y

CM

MY

CY CMY

K

10  

|

  QUESTIONS ABOUT TRIO? 1-888-335-6946

J

FUNCTIONALITY TEST

A Functionality Test  takes only a few minutes and is helpful if you
suspect the Trio Nebulizer Handset is not working properly. Follow the steps
below and use only 1 ml sodium chloride 0.9% (normal saline) for the
Functionality Test. (Please note that sodium chloride 0.9% is, by US Federal
law, restricted to sale by or on the order of a physician.)

1. Assemble the Trio Nebulizer Handset as described in Section F, but do not

attach the Mouthpiece.

2. Follow the procedure in Section G for taking a treatment. However, for

this Functionality Test only, use 1 ml of sodium chloride 0.9% instead
of your prescribed medication.

3. Using a stop watch or clock with a second hand, time the complete

Functionality Test from when the Trio Nebulizer Handset starts producing
aerosol until the Controller automatically turns off.

4. After the Controller has automatically shut off, remove the Medication

Cap and check the amount of liquid in the Medication Reservoir. Only
a few drops of saline should remain in the Medication Reservoir. If more
than a few drops remain, clean the Aerosol Head and repeat the test.
If, after proper cleaning, there are still more than a few drops remaining
in the Medication Reservoir, call the number below for details.

5. If the time to nebulize 1 ml of sodium chloride 0.9% is about 3 minutes,

the Trio Nebulizer Handset is functioning correctly. If it takes more than
4-5 minutes to nebulize 1 ml of liquid, call the number below for details.
Clean the Trio Nebulizer Handset after this Functionality Test, and prior
to using the your prescribed medication.

K  

FREQUENTLY ASKED QUESTIONS AND

TROUBLESHOOTING

QUESTION 1:

How much medication should be left in the Medication

Reservoir at the end of a treatment?

Answer:

 Only a few drops of medication should remain after the

treatment in the Medication Reservoir. If more than a few drops remain,
clean the Aerosol Head (Section H). If, after proper cleaning, more than
a few drops remain in the Medication Reservoir, call 1-888-335-6946.

QUESTION 2:

How long should each treatment take?

Answer:

  With normal operation and proper cleaning the Trio Nebulizer

Handset should deliver a 1 ml sodium chloride 0.9% treatment within
3 minutes. Perform a Functionality Test (Section J). If more than 4-5
minutes are required to complete your treatment with 1ml of sodium
chloride 0.9%, call 1-888-335-6946.

QUESTION 3:

When I turn on my Trio Electronic Nebulizer , nothing happens.

There does not seem to be any power.

Answer:

  If you are using batteries, use the figures in each battery

compartment to verify that the batteries are inserted correctly. If the
batteries are positioned properly, check to see if the light is blinking
on the Controller signaling low battery power. If it is, replace the batteries
or use the AC Power Supply.

Answer:

  If you are using the AC Power Supply, check the adapter

connection to be sure it is firmly connected to the adapter port located
underneath the battery compartment. Be sure that the adapter is
plugged into a working wall outlet.

Continued on back page

I

TRIO NEBULIZER HANDSET DISINFECTION

WARNING

To reduce the risk of serious or fatal illness caused by contamination of
the Trio Nebulizer Handset you must also DISINFECT THE HANDSET AT
THE END OF EVERY TREATMENT DAY. In addition, the Trio Nebulizer Handset
is strictly FOR SINGLE PATIENT USE ONLY.

1.  Clean the Trio Nebulizer Handset as described in Section H.

2. At the end of every treatment day, disinfect the Trio Nebulizer Handset

with a commercially available medical device disinfectant such as
Control III® (Figure 32). Follow the disinfectant manufacturer's
instructions for use.

3. After the Trio Nebulizer Handset parts are completely dry, place them

in a bag for storage. Re-assemble the Trio Nebulizer Handset just prior
to the next treatment. This is to ensure that the silicone seals do not
become deformed over time.

32

Figure

Содержание Trio

Страница 1: ...ral law restricts this devicetosalebyoronorderofaphysician Read and understand these Instructions for Use and all safety precautions it contains Improperusecancauseserious or fatal injury and or illne...

Страница 2: ...on and safety precautions for the Trio Electronic Nebulizer Save these Instructions For Use for future reference BeforeusingyourTrioElectronicNebulizer youshouldreadandunderstand the entire Instructio...

Страница 3: ...NEVER operate the Controller if it is improperly or incompletely assembled or damaged Section G Taking a treatment shows the alerts to these situations when the Trio Electronic Nebulizer is improperl...

Страница 4: ...o not have all the items please call the number listed below FOUR AA BATTERIES MEDICATION RESERVOIR AEROSOL CHAMBER MOUTHPIECE WITH EXHALATION VALVE AC POWER SUPPLY AEROSOL HEAD INHALATION VALVE CARRY...

Страница 5: ...eries will provide approximately 2 hours of continuous treatment You may also use AA rechargeablebatteries Pleasesee Specifications ontheback cover of these Instructions For Use NOTE IT IS STRONGLY RE...

Страница 6: ...fingers Figure 8 F TRIO NEBULIZER HANDSET ASSEMBLY WARNING TheTrioNebulizerHandsetisnotsterile CleanandDisinfecttheTrio NebulizerHandsetbeforethefirstuse seeSectionHforCleaningand SectionIforDisinfect...

Страница 7: ...withTrioElectronic Nebulizer a Trio Nebulizer Handset Attached To Controller Place the Controller on a clean flat stable surface Slide the Trio Nebulizer Handset into the Controller cradle as far as i...

Страница 8: ...en delivered a You will hear the Treatment Complete Alarm 2 beeps and 2 green lights b The Trio Electronic Nebulizer will automatically shut off 10 If for any reason you must stop the treatment before...

Страница 9: ...bulizerHandsetimmediatelyaftereveryuse Donot wash the Controller Nebulizer Connection Cord AC Power Supply or PlugConverters Useliquidsoapdesignedforwashingdishes Donot use white liquiddishsoaps e g I...

Страница 10: ...f clear liquid dish soap to a clean bowl or basin and add warm tap water Figure 28 5 Wash all parts of the Trio Nebulizer Handset and rinse thoroughly under running warmtapwater Figure 29 If Disinfect...

Страница 11: ...ain after the treatmentintheMedicationReservoir Ifmorethanafewdropsremain cleantheAerosolHead SectionH If afterpropercleaning morethan afewdropsremainintheMedicationReservoir call1 888 335 6946 QUESTI...

Страница 12: ...batteries following the procedure in Section D Please note that the Controller has a built in automatic shutoff after 10minutes anditwillturnoffatthattime Ifthissituationcontinues three or four times...

Отзывы: