DRYING
Even if the cleaner-disinfector is equipped with a drying func-
tion, make sure that no residual moisture is left in the compo-
nents. If necessary, remove any residual moisture by shaking
the parts and allow them to dry completely.
- 25 -
en
6.2 Manual cleaning
As an alternative to mechanical cleaning and disinfection, the
following manual method can be used:
EQUIPMENT
– Cleaning agent
– Brush
– Warm water
PROCEDURE
• Place all the parts in the cleaning solution.
• Clean the parts thoroughly with a brush.
• Thoroughly rinse the parts under hygienically tested water
and leave them to dry completely on a clean, dry and
absorbent surface.
VALIDATION
The method has been validated using:
– Korsolex
®
Endo Cleaner 0.5% (Bode)
– Approx. 50 °C warm water
– Duration of cleaning procedure: approx. 5 minutes